Inappropriate implantable cardioverter-defibrillator shocks attributed to alternating-current leak in a swimming pool
John N Makaryus1, Julia Angert-Gilman, Mena Yacoub
1Department of Cardiology, Division of Electrophysiology, North Shore-Long Island Jewish Health System, New Hyde Park, New York 11040.
Abstract:
Implantable cardioverter-defibrillators (ICDs) are the standard of care for preventing sudden cardiac death in patients who are predisposed to malignant ventricular arrhythmias. Causes of inappropriate ICD shock include equipment malfunction, improper arrhythmia evaluation, misinterpretation of myopotentials, and electromagnetic interference. As the number of implanted ICDs has increased, other contributors to inappropriate therapy have become known, such as minimal electrical current leaks that mimic ventricular fibrillation. We present the case of a 63-year-old man with a biventricular ICD who received 2 inappropriate shocks, probably attributable to alternating-current leaks in a swimming pool. In addition, we discuss ICD sensitivity and offer recommendations to avoid similar occurrences.