Impact of measurement parameters on apparent diffusion coefficient quantification in diffusion-weighted-magnetic
Holger Schmidt1, Sergios Gatidis, Nina F Schwenzer
1From the *Department of Radiology, Diagnostic and Interventional Radiology, and †Department of Radiology, Diagnostic and Interventional Radiology, Section on Experimental Radiology, Eberhard Karls University, Tübingen, Germany.
High reproducibility in apparent diffusion coefficient (ADC) measurements is achievable. Optimal b values and constant echo time (TE) are crucial for reliable ADC quantification across various tissues.
Area of Science:
- Medical Imaging
- Diffusion-Weighted Imaging (DWI)
- Quantitative MRI
Background:
- Apparent diffusion coefficient (ADC) is a key MRI biomarker.
- Reproducibility of ADC measurements is critical for clinical applications.
- Understanding the impact of imaging parameters on ADC is essential.
Purpose of the Study:
- To systematically evaluate the reproducibility of DWI.
- To assess the impact of b values and echo time (TE) on ADC quantification.
- To determine the minimum upper b value for reliable ADC measurements.
Main Methods:
- Phantom studies with varying diffusivities and T2 relaxation times.
- Diffusion-weighted imaging acquired with multiple b values and TEs.
- ADC calculated using various b value combinations; reproducibility assessed via coefficient of variation (CV) and CVsta.
- Volunteer studies evaluating ADC in 7 organs.
Main Results:
- ADC reproducibility improves with higher b values and increased diffusivity.
- An upper b value of 600 s/mm provides reliable ADC results (max CVsta 7.5%).
- ADC values decrease with increasing upper b value and TE; increase with T2 relaxation time.
Conclusions:
- ADC measurements exhibit high reproducibility but are sensitive to b values and TE.
- Constant TE is recommended for consistent ADC quantification.
- An upper b value of 600 s/mm offers a balance of ADC stability, SNR, and measurement time for diverse tissues.