Radiation safety consideration during intraoperative radiation therapy
Paul N Mobit1, Priyadarshini Rajaguru2, Michael Brewer2
1Department of Radiation Oncology, University of Mississippi Medical Center, 350 West Woodrow Wilson Drive, Suite 1600, Jackson, MS 39213, USA mobit_paul@yahoo.ca.
This study assessed radiation exposure from the Axxent Electronic Brachytherapy System during intraoperative radiation therapy (IORT). Movable lead shields significantly reduce exposure, allowing safe use in standard operating rooms with protective gear.
Area of Science:
- Medical Physics
- Radiation Oncology
Background:
- Intraoperative radiation therapy (IORT) systems require careful radiation safety assessments.
- The Axxent Electronic Brachytherapy System is a novel IORT device necessitating evaluation of its radiation output.
Purpose of the Study:
- To measure radiation exposure rates around the Axxent Electronic Brachytherapy System during phantom and clinical treatments.
- To determine the effectiveness of lead shielding in reducing radiation exposure.
- To assess the feasibility of using the system in standard operating rooms.
Main Methods:
- Utilized in-house-designed phantoms for initial radiation exposure measurements.
- Conducted radiation surveys during clinical IORT procedures on breast, floor of the mouth, and neck cancer patients.
- Measured exposure rates at various distances and assessed the impact of movable lead shields.
Main Results:
- Exposure rates of 200 mR/h at 30 cm were recorded during IORT with surface applicators and balloons.
- Movable lead shields reduced radiation exposure rates by over 95%.
- The 50-kVp X-ray source allows IORT in regular operating rooms with appropriate shielding and personal protective equipment.
Conclusions:
- The Axxent Electronic Brachytherapy System can be safely operated in standard operating rooms.
- Movable lead shields are highly effective in mitigating radiation exposure.
- Adherence to safety protocols, including lead aprons and shields, is crucial for personnel protection during IORT.
More Related Videos
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Aneurysm IV: Nursing Management
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...