Implementation of a dose gradient method into optimization of dose distribution in prostate cancer 3D-CRT plans
Marta K Giżyńska1, Paweł F Kukołowicz2, Paweł Kordowski3
1Department of Medical Physics, Maria Sklodowska-Curie Memorial Cancer Centre and Institute of Oncology, Warsaw, Poland ; Biomedical Physics Division, Institute of Experimental Physics, Faculty of Physics, University of Warsaw, Poland.
Aim:
The aim of this work is to present a method of beam weight and wedge angle optimization for patients with prostate cancer.
Background:
3D-CRT is usually realized with forward planning based on a trial and error method. Several authors have published a few methods of beam weight optimization applicable to the 3D-CRT. Still, none on these methods is in common use.
Materials And Methods:
Optimization is based on the assumption that the best plan is achieved if dose gradient at ICRU point is equal to zero. Our optimization algorithm requires beam quality index, depth of maximum dose, profiles of wedged fields and maximum dose to femoral heads. The method was tested for 10 patients with prostate cancer, treated with the 3-field technique. Optimized plans were compared with plans prepared by 12 experienced planners. Dose standard deviation in target volume, and minimum and maximum doses were analyzed.
Results:
The quality of plans obtained with the proposed optimization algorithms was comparable to that prepared by experienced planners. Mean difference in target dose standard deviation was 0.1% in favor of the plans prepared by planners for optimization of beam weights and wedge angles. Introducing a correction factor for patient body outline for dose gradient at ICRU point improved dose distribution homogeneity. On average, a 0.1% lower standard deviation was achieved with the optimization algorithm. No significant difference in mean dose-volume histogram for the rectum was observed.
Conclusions:
Optimization shortens very much time planning. The average planning time was 5 min and less than a minute for forward and computer optimization, respectively.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Dose Response Curve: Conventional Versus Nonmonotonic
Dosage Regimens: Designs and Approaches
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations