Related Experiment Video
Updated: Apr 21, 2026

Dissolved Solute Sampling Across an Oxic-Anoxic Soil-Water Interface Using Microdialysis Profilers
Published on: March 24, 2023
A flow-injection system coupled to a micro-guard cartridge for monitoring a vinification process
Susana S M P Vidigal1, António O S S Rangel
1CBQF-Centro de Biotecnologia e Química Fina-Laboratório Associado, Escola Superior de Biotecnologia, Universidade Católica Portuguesa/Porto.
Abstract:
The present paper describes a flow-injection system coupled to a Micro-Guard cartridge and a miniaturized optical CCD detection system used to monitor the sugars (glucose/fructose) and ethanol content during alcoholic fermentation. The carrier stream (mobile phase) is composed of an aqueous sulfuric acid solution. The flow parameters were studied in order to obtain a good resolution and a wide dynamic concentration range, with a good repeatability. The relative standard deviation (RSD) obtained was inferior to 4%, for n = 3. It was possible to achieve a linear range of up to 12 g L(-1) of sugars and up to 2% (v/v) of ethanol with a detection limit of 2.3 g L(-1) and 0.4% (v/v), for sugars and ethanol concentrations, respectively. The proposed system was successfully applied to monitor a vinification process by the quantification of sugars and ethanol, and also in some finished port wines.
Related Concept Videos
Dialysis
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...