Effect of tube current modulation for dose estimation using a simulation tool on body CT examination
Ai Kawaguchi1, Yuta Matsunaga2, Masanao Kobayashi3
1Department of Radiology, Toyota Memorial Hospital, Aichi 471-0821, Japan Department of Radiological Technology, Graduate School of Medicine, Tohoku University, 2-1 Seiryo-machi, Aoba-ku, Sendai, Miyagi 980-8575, Japan hanamaru0918@gmail.com.
Tube current modulation in body CT affects organ dose estimation, with bladder dose increasing and breast dose decreasing. Longitudinal variations were similar between iterative and filtered back-projection reconstruction algorithms.
Area of Science:
- Medical Imaging
- Radiological Physics
- Radiation Dosimetry
Background:
- Accurate dose estimation in computed tomography (CT) is crucial for patient safety.
- Tube current modulation (TCM) is a technique used to optimize radiation dose in CT examinations.
- Understanding the impact of TCM on organ dose is essential for precise risk assessment.
Purpose of the Study:
- To evaluate the impact of tube current modulation on dose estimation in body CT.
- To compare longitudinal variations in tube current values between iterative reconstruction (IR) and filtered back-projection (FBP) algorithms.
- To assess the differences in organ and effective doses using average versus modulated tube current values.
Main Methods:
- Simulation tool used for dose estimation in 100 body CT patients.
- Longitudinal tube current values recorded around 10 organ regions.
- Organ and effective doses simulated using average and modulated tube current values.
Main Results:
- Organ doses for bladder (20% higher) and breast (25% lower) showed significant differences with modulated TCM compared to average values.
- Differences in estimated effective doses were minimal (mean 0.7 mSv).
- Longitudinal tube current variations were comparable between IR and FBP algorithms.
Conclusions:
- Tube current modulation significantly impacts specific organ dose estimations in body CT.
- The choice between IR and FBP algorithms has minimal effect on longitudinal tube current variations.
- Accurate dose assessment requires considering TCM effects on individual organs.