Increased perforation risk with an MRI-conditional pacing lead: a single-center study.
Moshe Rav Acha1, John J Keaney, Steven A Lubitz
1Cardiac Arrhythmia Service, Massachusetts General Hospital, Boston, Massachusetts.
This study found higher perforation rates in patients with Medtronic MRI SureScan leads compared to non-MRI leads. Delayed perforations were common, suggesting a need for further investigation into MRI SureScan lead safety.
Area of Science:
- Cardiovascular Medicine
- Medical Device Technology
- Radiology
Background:
- Cardiac pacemakers (PPMs) were historically contraindicated for Magnetic Resonance Imaging (MRI).
- The introduction of Medtronic (MDT) MRI SureScan PPM and leads aimed to address this limitation.
- This study evaluates the safety of these newer MRI-compatible pacemaker leads.
Purpose of the Study:
- To compare lead perforation and dislodgement rates between MDT CapSureFix 5086 MRI SureScan leads and MDT non-MRI SureScan active fixation leads.
- To assess the safety profile of MRI-compatible pacemaker leads in clinical practice.
Main Methods:
- Retrospective analysis of patient records from April 2011 to April 2014.
- Inclusion of patients implanted with MDT CapSureFix 5086 MRI SureScan leads and MDT 4076/5076 non-MRI leads.
- Comparison of lead perforation and dislodgement events between the two lead types.
Main Results:
- A significantly higher perforation rate (5.5%) was observed with MDT CapSureFix 5086 MRI SureScan leads compared to non-MRI leads (0.47%) (P = 0.005).
- Three of the four perforations with MRI SureScan leads were delayed, occurring over 3 weeks post-implantation.
- A trend towards increased dislodgement rates was noted for MRI SureScan leads, though not statistically significant (P = 0.18).
Conclusions:
- This retrospective study suggests an increased risk of lead perforation with MDT CapSureFix 5086 MRI SureScan leads, contrary to some prior research.
- Delayed perforations represent a significant concern associated with these MRI-compatible leads.
- Further large-scale, prospective studies with extended follow-up are warranted to validate these findings and confirm lead safety.