Optimizing contrast media injection protocols in state-of-the art computed tomographic angiography
Michael M Lell1, Gregor Jost, Johannes Georg Korporaal
1From the *Department of Radiology, University of Erlangen; †Imaging Science Institute Erlangen, Erlangen; ‡MR and CT Contrast Media Research, Bayer Healthcare, Berlin; §Siemens Healthcare, Forchheim; and ∥Department of Radiology, University of Marburg, Marburg, Germany.
Objective:
Very short acquisition times and the increasing use of low-kilovolt protocols in standard computed tomographic (CT) angiography (CTA) examinations demand modifications in the contrast media (CM) injection regimen. The aim of this study was to optimize the use of tube voltage-adjusted CM delivery parameters, especially injection duration and iodine delivery rate (IDR), in thoracoabdominal CTA in a porcine model.
Materials And Methods:
Eight pigs (53-72 kg) were examined with a third-generation dual-source CT system with a dynamic CTA protocol (4-dimensional spiral, 454-mm scan length, 2.5-second temporal resolution, 70-second total acquisition time). Six CM injection protocols were applied in randomized order and intraindividually compared. The standard CTA protocol was performed at 120 kV, with an injection of 300 mg iodine/kg body weight and a flow of 5 mL/s (IDR, 1.5 g/s). On the basis of phantom measurements for the low-kilovolt CTA protocols, the iodine dosage was adjusted to 150 mg iodine/kg (70 kV) and 210 mg iodine/kg (90 kV). Therefore, either the IDR was kept constant and the injection time was reduced, or the injection time was kept constant and the IDR was reduced by modifying the CM flow or concentration. Time attenuation curves, time to peak, and peak enhancement were calculated for different locations within the aorta, renal arteries, and large veins.
Results:
The heart rates were comparable among the different injection protocols (66.9-78.1 beats per minute). The average injection peak pressure depended on the flow rate and CM concentration and ranged from 42.9 to 114.7 psi. The average arterial peak enhancement was comparable for protocols with identical injection times and reduced IDR (362.4 HU [standard 120-kV protocol; 300 mg iodine/kg; IDR, 1.5 g/s], 360.0 HU [70 kV; 150 mg iodine/kg; IDR, 0.75 g/s], 365.4 HU [70 kV; 150 mg iodine/kg; IDR, 0.75 g/s; CM, 150 mg iodine/mL], 344.3 HU [90 kV; 210 mg iodine/kg; IDR, 1.1 g/s]). Higher peak enhancements could be achieved by applying protocols with identical IDR and a reduced injection time (502.5 HU [70 kV; 150 mg iodine/kg; IDR, 1.5 g/s] and 394.6 HU [90 kV; 210 mg iodine/kg; IDR, 1.5 g/s]).
Conclusions:
By adjusting the IDR, low-kilovolt CTA is able to achieve comparable aortic enhancement with a significant reduction in CM dosage. A shorter injection time at constant IDR results in higher enhancement and a narrower scan window and might be preferable for fast CTA acquisition techniques.
Clinical Relevance/Application:
The optimization of CM injection protocols is mandatory to achieve state-of the art CTA at low kilovolt and can reduce CM doses to patients.
More Related Videos
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
10:26A Multicenter MRI Protocol for the Evaluation and Quantification of Deep Vein Thrombosis
Published on: June 2, 2015
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...