Multi-centre variability of ISCEV standard ERGs in two normal adults
R Hamilton1, A Al Abdlseaed, J Healey
1Department of Clinical Physics, Royal Hospital for Sick Children, NHS Greater Glasgow and Clyde, University of Glasgow, Dalnair Street, Glasgow, G3 8SJ, UK, ruth.hamilton@glasgow.ac.uk.
Documenta Ophthalmologica. Advances in Ophthalmology
|February 28, 2015
Summary
Reproducibility of electroretinograms (ERGs) was assessed across 15 UK centers. Results show minimal inter-center variability, supporting the effectiveness of ERG standardization for reliable clinical use.
Area of Science:
- Ophthalmology
- Neuroscience
- Biomedical Engineering
Background:
- Reproducibility of full-field electroretinograms (ERGs) between different clinical centers has not been previously evaluated.
- Assessing inter-center variability is crucial for establishing reliable diagnostic standards in ophthalmology.
Purpose of the Study:
- To investigate the reproducibility of full-field ERGs across multiple centers.
- To compare the variability of ERGs recorded using local versus a standardized ISCEV (International Society for Clinical Electrophysiology of Vision) protocol.
- To determine if standardization enhances ERG reliability.
Main Methods:
- Electroretinograms (ERGs) were recorded from two subjects at 15 UK centers using both silver thread and skin electrodes.
- Recordings were performed using routine local protocols and a standardized ISCEV-specified protocol.
- Variability was assessed for amplitude and peak time across centers and protocols.
Main Results:
- Inter-center ERG variability was generally small and within clinically significant thresholds.
- Amplitude variability was lower with local LED protocols compared to the ISCEV-specified xenon flash protocol.
- Peak time variability was lower with the ISCEV-specified protocol, suggesting improved standardization benefits.
Conclusions:
- Inter-center variability in ERGs is often within acceptable clinical limits, especially with appropriate standardization.
- The existing ERG standard is vindicated, and further standardization could improve global ERG reproducibility.
- Standardization efforts should consider the impact of equipment like xenon flashtubes on overall variability.