Related Experiment Video
Updated: Apr 15, 2026

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Accuracy of a dose map method assessed in clinical and anthropomorphic phantom situations using Gafchromic films
C Bordier1, R Klausz2, L Desponds2
1Detection and Guidance Solutions, Image Quality and Dose Center of Excellence, GE Healthcare, 283 rue de la Minière, Buc 78530, France clemence.bordier@ge.com.
Abstract:
A dose map method has been integrated on GE x-ray angiographic systems to provide an indication of the local dose distributed on a patient envelope representative of individual patient shapes. Tests have been performed to assess the accuracy of the method by using Gafchromic XR-RV3 films in an anthropomorphic phantom situation and in a clinical situation. Dose values inside different exposed areas have been compared between the film and the dose map method. The dose map results show a good visual agreement for the anthropomorphic phantom situation, and the local doses agreed within better than 15 % compared with the Gafchromic films in both situations.

