Ultrafast cone-beam computed tomography: a comparative study of imaging protocols during image-guided therapy
Jijo Paul1, Annamma Chacko2, Mohammad Farhang1
1Diagnostic and Interventional Radiology, University Hospital, Goethe University Frankfurt, Theodor-Stern-Kai 7, 60590 Frankfurt, Germany.
Objective:
To evaluate two ultrafast cone-beam CT (UF-CBCT) imaging protocols with different acquisition and injection parameters regarding image quality and required contrast media during image-guided hepatic transarterial chemoembolization (TACE).
Methods:
In 80 patients (male: 46, female: 34; mean age: 56.8 years; range: 33-83) UF-CBCT was performed during TACE for intraprocedural guidance. Imaging was performed using two ultrafast CBCT acquisition protocols with different acquisition and injection parameters (imaging protocol 1: acquisition time 2.54 s, and contrast 6 mL with 3 s delay; imaging protocol 2: acquisition time 2.72 s, and contrast 7 mL with 6 s delay). Image evaluation was performed with both qualitative and quantitative methods. Contrast injection volume and dose parameters were compared using values from the literature.
Results:
Imaging protocol 2 provided significantly better (P < 0.05) image quality than protocol 1 at the cost of slightly higher contrast load and patient dose. Imaging protocol 1 provided good contrast perfusion but it mostly failed to delineate the tumors (P < 0.05). On the contrary, imaging protocol 2 showed excellent enhancement of hepatic parenchyma, tumor, and feeding vessels.
Conclusion:
Tumor delineation, visualization of hepatic parenchyma, and feeding vessels are clearly possible using imaging protocol 2 with ultrafast CBCT imaging. A reduction of required contrast volume and patient dose were achieved due to the ultrafast CBCT imaging.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET