Related Experiment Video
Updated: Apr 13, 2026

Microfluidics in Assessing Platelet Function
Published on: November 8, 2024
Assessment of platelet function in healthy sedated cats using three whole blood platelet function tests
Kimberly K Ho1, Anthony C G Abrams-Ogg2, R Darren Wood1
1Departments of Clinical Studies (Ho, Abrams-Ogg, O'Sullivan, Blois)Pathobiology (Wood), Ontario Veterinary College, University of Guelph, Guelph, Ontario, CanadaBiomedical Sciences (Kirby), Ontario Veterinary College, University of Guelph, Guelph, Ontario, Canada.
Abstract:
The objectives of this study were to establish feline references intervals for 3 commercial whole blood platelet function test analyzer systems: Multiplate analyzer (MP; Roche Diagnostics International Ltd., Rotkreuz, Switzerland), Platelet Function Analyzer-100 (PF: Siemens Canada, Mississauga, Ontario, Canada), and Plateletworks Combo-25 kit (PW; Helena Laboratories, Beaumont, TX). Venipuncture was performed on 55 healthy sedated cats, and platelet aggregation in response to adenosine diphosphate (ADP), collagen (COL), and arachidonic acid (AA; MP only) was assessed using citrated blood. For the MP analyzer, median (95% confidence intervals [CIs]) area under curve (Units) for ADP, COL, and AA agonists were 87 (11-176), 81 (32-129), and 91 (59-129), respectively. For the PF analyzer, median (95% CIs) closure time, using COL-ADP cartridges, was 69 (46-89) sec. For the PW assay, median (95% CIs) percent aggregations for ADP and COL agonists were 71 (18-92) and 49 (9-96), respectively, using impedance hematology analyzer platelet counts, and 94 (25-98) and 68 (14-119), respectively, using flow cytometry hematology analyzer platelet counts. There were low correlations between the PF analyzer (COL-ADP cartridge) and MP analyzer (COL agonist; ρ = 0.11), and between the PF analyzer (COL-ADP cartridge) and PW assay (COL agonist using impedance platelet counts; ρ = 0.14). The PW assay percent aggregations using impedance and flow cytometric platelet counts were correlated for both ADP (ρ = 0.64) and COL (ρ = 0.64) agonists. Platelet function testing using these tests are feasible in cats, but 95% CIs are wide, so single results may be difficult to interpret. Platelet counting by impedance or flow cytometry may be used for the PW assay but are not interchangeable.

