Background:
The safety and efficacy of exchanging the HeartMate II (Thoratec Corp, Pleasanton, CA) left ventricular assist device (LVAD) through a less invasive subcostal approach remains unclear.
Methods:
We reviewed the records of 17 patients who underwent exchange of their HeartMate II device at our institution since 2007. We divided the cohort into devices exchanged through a subcostal (SC) approach versus a median sternotomy (MS) approach and obtained data pertaining to the short- and long-term outcomes.
Results:
Nine patients had pump exchange through an MS approach versus 8 patients who underwent an SC approach. The mean duration of support with the first pump was 540 ± 450 days. The reason for exchange was electromechanical failure (7 patients), thrombosis (8 patients), and infection (2 patients). There were no 30-day perioperative deaths with either approach. Compared with sternotomy, patients who underwent an SC approach had significantly shorter operative times (131 vs 222 minutes; p = 0.001) and lower reoperation rates for bleeding (0 vs 44.4%; P = 0.05) and required fewer transfused blood products (packed red cells, 3.5 units vs 7.1 units; p < 0.05; cryoprecipitate, 50.7 mL vs 209.3 mL; p = 0.01; and platelets, 292 mL versus 762 mL; p < 0.05). Additionally, patients who underwent an SC approach had shorter postoperative stays in the intensive care unit (ICU) (5 days vs 13.8 days; p < 0.05) and shorter total hospital stays (16.4 days vs 27.2 days; p < 0.05). Long-term survival after mean follow-up of 260 days for the SC group and 232 days for the sternotomy group was 75% and 33%, respectively.
Conclusions:
Exchange of the HeartMate II pump can be accomplished with low morbidity and mortality and good long-term outcomes through a less invasive SC approach.