2240 IRRADIATED PERSONAL DOSEMETERS FOR THE IC2014 EURADOS INTERCOMPARISON AT SEIBERSDORF
1Dosimetry Laboratory Seibersdorf, Seibersdorf Labor GmbH, Seibersdorf 2444, Austria christian.hranitzky@seibersdorf-laboratories.at.
Radiation Protection Dosimetry
|August 28, 2015
Summary
The Dosimetry Laboratory Seibersdorf evaluated 2240 personal dosemeters for the EURADOS intercomparison, assessing their accuracy in measuring personal dose equivalent (Hp(10) and Hp(0.07)) across various photon radiation fields.
Area of Science:
- Medical Physics
- Radiation Dosimetry
- Nuclear Engineering
Background:
- The IC2014 EURADOS intercomparison is crucial for validating individual monitoring systems.
- Accurate dosimetry is essential for radiation protection and regulatory compliance.
- The Dosimetry Laboratory Seibersdorf has a key role in such intercomparisons.
Purpose of the Study:
- To serve as the irradiation laboratory for the IC2014 EURADOS intercomparison.
- To irradiate a large number of personal dosemeters under controlled conditions.
- To assess the performance of whole-body dosimetry systems in terms of personal dose equivalent.
Main Methods:
- Irradiation of 2240 personal dosemeters from 112 dosimetry systems.
- Use of ISO water slab phantom for simulating human exposure.
- Irradiation with photon reference fields including Cs-137, Co-60, and 320-kV X-rays.
- Employing nine irradiation setups with five ISO/IEC photon radiation qualities and two angles of incidence (0° and 60°).
Main Results:
- Reference dose equivalent values ranged from approximately 1 to 500 mSv.
- Reference values were established using air kerma rate traceable to primary standards.
- Total expanded measurement uncertainties for reference values were less than 6%.
Conclusions:
- The Dosimetry Laboratory Seibersdorf successfully conducted the irradiation phase of the IC2014 EURADOS intercomparison.
- The established reference dosimetry provided a reliable basis for evaluating participating dosimetry systems.
- The results contribute to the overall quality assurance of individual monitoring in Europe.