Note: Design and realization of a portable edge illumination X-ray phase contrast imaging system
D Basta1, M Endrizzi1, F A Vittoria1
1Department of Medical Physics and Bioengineering, University College London, Malet Place, Gower Street, London WC1E 6BT, United Kingdom.
Abstract:
We discuss a portable edge illumination x-ray phase contrast imaging system based on compact piezoelectric motors, which enables its transportation to different environments, e.g., hosting different x-ray source technologies. The analysis of images of standard samples reveals an angular sensitivity of 270 ± 6 nrad, which compares well with the 260 ± 10 nrad reported for previous systems based on stepper motors, demonstrating that system portability can be achieved without affecting phase sensitivity. The results can also be considered a test of the performance of the piezoelectric motors, and as such could be of interest to researchers planning their use in other imaging systems.
Related Concept Videos
X-ray Imaging
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...