Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
A pencil beam approach to proton computed tomography.
Regina Rescigno1, Cécile Bopp1, Marc Rousseau1
1Université de Strasbourg, IPHC, 23 rue du Loess, Strasbourg 67037, France and CNRS, UMR7178, Strasbourg 67037, France.
A novel beam-by-beam approach for proton computed tomography (pCT) redefines reconstruction elements and proposes an analytical algorithm. This method may overcome challenges of the current proton-by-proton imaging technique.
Area of Science:
- Medical Imaging
- Particle Physics
- Computational Science
Background:
- Proton computed tomography (pCT) offers superior soft-tissue contrast compared to X-ray CT.
- Current pCT methods track protons individually, posing technical challenges.
- A new paradigm considering proton beams rather than individual particles is explored.
Purpose of the Study:
- To introduce a novel beam-by-beam approach for pCT.
- To redefine pCT reconstruction elements (residual energy, path) for this new approach.
- To propose an analytical image reconstruction algorithm for the beam-by-beam method.
Main Methods:
- Modeled proton beam (PB) propagation using generalized Fermi-Eyges theory with multiple Coulomb scattering (MCS).
- Integrated the PB model into pCT reconstruction, defining a mean beam path.
- Adapted filtered backprojection algorithm for the beam-by-beam approach and simulated image reconstruction.
Main Results:
- The PB model showed good agreement for mean values but depth-dependent discrepancies (2-8%) in distribution spread compared to Monte Carlo simulations.
- Images reconstructed with the beam-by-beam approach exhibited higher noise than those from the classical single-proton method.
- The study validated the analytical reconstruction algorithm's capability to incorporate MCS.
Conclusions:
- The beam-by-beam approach offers a potential solution to technical limitations of current pCT methods.
- An analytical algorithm is presented for this new approach, accounting for MCS during reconstruction.
- Further research is needed to fully assess the image quality and limitations of the beam-by-beam pCT method.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Imaging Studies III: Computed Tomography