A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
K2-EDTA and K3-EDTA Greiner Tubes for HbA1c Measurement
Alen Vrtaric1, Petra Filipi2, Marina Hemar2
1University Department of Chemistry, Medical School University Hospital Sestre Milosrdnice, Zagreb, Croatia alenvrtaric@gmail.com.
Objective:
To determine whether K2-ethylenediaminetetraacetic acid (EDTA) and K3-EDTA Greiner tubes could be used interchangeably for glycosylated hemoglobin, type A1C (HbA1c) measurement via the Abbott Laboratories ARCHITECT chemiluminescent microparticle HbA1c assay on the ARCHITECT i2000SR immunoanalyzer at our university hospital.
Materials:
We drew blood from a total of 45 outpatients into plastic Greiner Vacuette tubes, some of which were lined with K2-EDTA and others with K3-EDTA anticoagulant. Data are presented as median and interquartile range values. We used the Wilcoxon test and Passing-Bablok regression for tube comparison.
Results:
For K2-EDTA tubes median HbA1c concentration was 54 mmol/mol (41 to 71 mmol/mol) and for K3-EDTA tubes 56 mmol/mol (43 to 69 mmol/mol). There was no statistically significant difference between K2-EDTA and K3-EDTA (bias= -1.29 mmol/mol; P = 0.24). Passing-Bablok regression showed that there is no constant and proportional error: y = -0.23 (95% CI[-3.52 to 0.69]) + 1.00( 95% CI[0.98 to 1.06]) x.
Conclusion:
In this study, we provide evidence for the lack of any clinically and statistically significant bias between K2-EDTA and K3-EDTA HbA1c measurements. Thus, Greiner tubes lined with K2-EDTA and those lined with K3-EDTA can safely be used interchangeably to measure HbA1c via the Abbott Laboratories ARCHITECT assay.
More Related Videos
09:41Hemocompatibility Testing of Blood-Contacting Implants in a Flow Loop Model Mimicking Human Blood Flow
Published on: March 5, 2020
08:44An In Vitro Hemodynamic Loop Model to Investigate the Hemocytocompatibility and Host Cell Activation of Vascular Medical Devices
Published on: August 21, 2020
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
EDTA: Auxiliary Complexing Reagents
Complexometric EDTA Titration Curves
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...