Automated 90Sr Separation and Preconcentration in a Lab-on-Valve System at Ppq Level
Published on: June 6, 2018
Preparation of gaseous CRMs from the primary system for (222)Rn activity measurement
1Korea Research Institute of Standards and Science, Daejeon 305-340, Republic of Korea; University of Science & Technology (UST), Republic of Korea.
Abstract:
For disseminating the gaseous radon standard traceable to the KRISS primary system based on the defined solid angle counting method, two kinds of radon CRM (a glass ampule type and a stainless steel cylinder type) were developed. The activity of the CRM was certified by subtracting a residual activity from the measured activity by the primary system. After certification, the ampule CRM was used to calibrate a radon-monitoring instrument and the cylinder CRM to calibrate an HPGe system. We also improved the measurement procedure of the radon primary system. In a typical radon energy spectrum, the radon peak overlaps with the polonium peak. For more reliable and accurate measurement of radon activity, a fitting method was adopted for the evaluation of radon area in the alpha energy spectrum. The result of radon activity evaluated by using the fitting method is in good agreement with that by the previous integration method.