Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Measurement agreement between a new biometer based on partial coherence interferometry and a validated biometer based
Junhua Li1, Hao Chen1, Giacomo Savini1
1From the School of Ophthalmology and Optometry (Li, Chen, Lu, Yu, Bao, Wang, Huang), Wenzhou Medical University, Wenzhou, Zhejiang, and the Key Laboratory of Vision Science (Li, Chen, Lu, Yu, Bao, Wang, Huang), Ministry of Health, P.R. China, Wenzhou, Zhejiang, China G.B. Bietti Foundation IRCCS (Savini), Rome, Italy.
The AL-Scan optical biometer shows excellent agreement with the Lenstar for ocular measurements like axial length (AL) and central corneal thickness (CCT). However, pupil diameter measurements differed significantly between the two devices.
Area of Science:
- Ophthalmology
- Medical Devices
- Biometry
Background:
- Accurate ocular measurements are crucial for cataract surgery planning.
- Comparing new biometry devices to established ones ensures reliability.
- This study evaluates a new partial coherence interferometry (PCI) device against an optical low-coherence reflectometry (OLCR) device.
Purpose of the Study:
- To assess the agreement between a new optical biometer (AL-Scan, PCI) and a validated optical biometer (Lenstar, OLCR).
- To compare key ocular measurements including axial length, corneal thickness, and anterior chamber depth.
Main Methods:
- Prospective observational cross-sectional study involving 92 eyes of cataract patients.
- Measurements of axial length (AL), central corneal thickness (CCT), anterior chamber depth (ACD), keratometry (K), astigmatism, white-to-white (WTW), and pupil diameter were taken three times with each device.
- Bland-Altman analysis was used to assess agreement and limits of agreement (LoA).
Main Results:
- Excellent agreement (narrow 95% LoA) was found between the PCI and OLCR devices for AL, CCT, ACD, and aqueous depth.
- Good agreement was observed for mean K, astigmatism, and WTW, although statistically significant differences were noted.
- A significant discrepancy was found in pupil diameter measurements, with wide 95% LoA.
Conclusions:
- The AL-Scan (PCI) biometer provides ocular measurements comparable to the Lenstar (OLCR) for most parameters, particularly AL, CCT, and ACD.
- Pupil diameter measurements from the AL-Scan device showed poor agreement with the Lenstar and require careful interpretation and potential adjustment.