Quantitating Fluorescence Intensity From Fluorophores: Practical Use of MESF Values
Lili Wang1, Adolfas K Gaigalas1, Fatima Abbasi2
1National Institute of Standards and Technology, Gaithersburg, MD 20899-8312.
Summary
Accurate flow cytometry requires matching emission spectra and normalizing by extinction coefficient. Fluorescein pH is critical, and a specific microbead (bead7) shows potential as a superior calibration standard.
Area of Science:
- Biophysics
- Analytical Chemistry
- Cell Biology
Background:
- Quantitative flow cytometry relies on accurate fluorescence measurements.
- Molecules of Equivalent Soluble Fluorophores (MESF) are used for calibration.
- Standardization of calibration methods is essential for reliable results.
Purpose of the Study:
- To identify critical parameters for quantitative flow cytometry using MESF values.
- To evaluate fluorescein as a model fluorophore for calibration.
- To assess potential improvements in calibration standards for flow cytometry.
Main Methods:
- Utilized fluorescein as a model fluorophore.
- Investigated emission spectrum matching between reference solutions and analytes.
- Normalized fluorescence signals by extinction coefficients.
- Assessed the impact of pH on MESF assignments.
- Compared emission spectra of microbeads and biological cells.
Main Results:
- Emission spectrum matching and extinction coefficient normalization are crucial for accurate fluorescence quantification.
- The pH of the medium significantly affects MESF assignments when using fluorescein.
- Microbead suspensions and stained biological cells exhibit similar emission spectrum shapes.
- A microbead with a seven-methylene linker (bead7) demonstrated the best spectral match with biological cells.
Conclusions:
- Optimized protocols for MESF-based flow cytometry require careful attention to spectral properties and environmental factors.
- The bead7 microbead shows promise as an improved calibration standard due to superior spectral compatibility with biological samples.
- Further validation of bead7 as a calibration standard could enhance the accuracy and reproducibility of flow cytometric analyses.
Related Concept Videos
Fluorescence and Phosphorescence: Instrumentation
1.8K
Fluorometers and spectrofluorometers are two types of instruments used for measuring molecular fluorescence. These instruments differ in how they select excitation and emission wavelengths and the type of light sources they utilize. Fluorometers use absorption interference filters to choose excitation and emission wavelengths. The excitation source in a fluorometer is typically a low-pressure mercury vapor lamp that emits intense lines distributed throughout the ultraviolet and visible regions.
1.8K
Real Time RT-PCR
66.3K
Real-time reverse transcription-polymerase chain reaction, or Real-time RT-PCR, is an analytical tool used to determine the expression level of target genes. The method involves converting mRNA to complementary DNA with the help of an enzyme known as reverse transcriptase, followed by the PCR amplification of the cDNA. These two processes can be performed simultaneously in a single tube or separately as a two-step reaction.
The real-time quantification of the number of amplified products is...
The real-time quantification of the number of amplified products is...
66.3K