Simultaneous ex vivo Functional Testing of Two Retinas by in vivo Electroretinogram System
Published on: May 6, 2015
Photopic Negative Response Obtained Using a Handheld Electroretinogram Device: Determining the Optimal Measure and
Zhichao Wu1, Xavier Hadoux1, Flora Hui1
1Centre for Eye Research Australia, Royal Victorian Eye and Ear Hospital, East Melbourne, Australia ; Ophthalmology, Department of Surgery, The University of Melbourne, Melbourne, Australia.
The photopic negative response/b-wave peak ratio (PhNR/B) offers the most repeatable full-field electroretinogram (ERG) measurement. This handheld ERG system with skin electrodes provides consistent results across various conditions for clinical research.
Area of Science:
- Ophthalmology
- Neuroscience
- Medical Devices
Background:
- The full-field electroretinogram (ERG) is crucial for assessing retinal function.
- The photopic negative response (PhNR) is a key component of the ERG, reflecting inner retinal function.
- Optimizing PhNR measurement repeatability is essential for reliable clinical and research applications.
Purpose of the Study:
- To identify the optimal measure of the photopic negative response (PhNR) for test-retest repeatability.
- To evaluate the repeatability of PhNR measurements using a handheld, nonmydriatic ERG system with self-adhering skin electrodes.
- To assess PhNR repeatability under varying conditions, including different examiners and sessions.
Main Methods:
- Twenty normal participants underwent multiple ERG recordings (200 sweeps each) in both eyes across two sessions.
- The coefficient of repeatability (CoR) was calculated for different PhNR measures and testing conditions.
- Comparisons were made for intra-examiner, inter-examiner, and inter-session variability.
Main Results:
- The PhNR/b-wave peak ratio (PhNR/B) demonstrated the lowest CoR relative to its dynamic range (30 ± 4%).
- No significant changes in the PhNR/B ratio were observed over seven measurements within sessions (P ≥ 0.100).
- No significant differences in CoR were found across different testing conditions (P = 0.314).
Conclusions:
- The PhNR/B ratio is the most reliable measure for minimizing variability in ERG recordings.
- The handheld ERG system and proposed protocols ensure consistent PhNR measurements.
- This technique is suitable for clinical and research use due to its high consistency.