Related Experiment Video
Updated: Mar 12, 2026

Using a Real-Time Locating System to Measure Walking Activity Associated with Wandering Behaviors Among Institutionalized Older Adults
Published on: February 8, 2019
A Community Hospital's Evaluation of Alarm Management Safety Factors
Emma Kurnat-Thoma1, Kayuri Shah
1Author Affiliations: Research Coordinator (Dr Kurnat-Thoma), Research Department, Reston Hospital Center, Reston; and Biomedical Engineering Student (Ms Shah), Department of Biomedical Engineering, Virginia Commonwealth University, Richmond, Virginia.
Background:
The Joint Commission's 2014 National Patient Safety Goals required hospitals to evaluate alarm safety in 2014-2015 and implement alarm safety policies.
Objective:
The aim of this study was to assess common alarm management safety factors in our 187-bed community hospital.
Methods:
Two weeks' worth of IV pump report data was evaluated to characterize 33 IV pump alarm types. Hospital and IV pump noise was measured, and an alarm management nurse survey was conducted.
Results:
There were 8731 total IV pump alarms/alerts (24-hour mean, 623.6) across 6 units. The 2-minute idle alarm accounted for 32.4% of all total IV alarms/alerts, suggestive of high levels of nurse multitasking and nurse work interruptions. IV pump volumes contributed to overall hospital noise. Survey data identified patient units and alarm safety practices needing additional support.
Conclusions:
Characterization of IV pump alarms/alerts is an emerging area of scientific inquiry. Findings indicate the need for organizations to evaluate alarm burden and alarm management safety practices to reduce alarm fatigue risks.
More Related Videos
Related Concept Videos
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Guidelines and Strategies for Safe Computer Charting
Maintain Confidentiality and Security:
Types of Reports II: Incident or Occurrence Report
Purposes:
In the healthcare industry, reports play a crucial role in documenting incidents within an agency. The primary objective of these reports is to ensure patient safety, uphold the...
Planning Nursing Care I
Acute Coronary Syndrome V: Nursing Management