Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
CoreValve Prosthesis Depth: What is the Optimal Measurement Target?
Rafael Wolff1,2, Sam Radhakrishnan1,3, Hirotsugu Mitsuhashi1,3
1Schulich Heart Program, Sunnybrook Health Sciences Centre, Toronto, Canada|.
Improper implantation depth of the Medtronic CoreValve (MCV) during transcatheter aortic valve replacement (TAVR) increases risks of heart conduction problems and permanent pacemaker (PPM) need. Angiographic measurements of prosthesis depth are key predictors for these complications.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Devices
Background:
- Transcatheter aortic valve replacement (TAVR) using the Medtronic CoreValve (MCV) is associated with significant conduction disturbances.
- Permanent pacemaker (PPM) implantation is a common complication following MCV TAVR.
- Prosthesis implantation depth is a potential factor influencing these adverse events.
Purpose of the Study:
- To investigate the relationship between MCV prosthesis implantation depth and the incidence of conduction disturbances.
- To determine if angiographic measurements of prosthesis depth predict the need for PPM after TAVR.
Main Methods:
- Retrospective analysis of 157 patients undergoing MCV TAVR (2009-2013).
- Exclusion of patients with pre-existing pacemakers (n=27).
- Definition of prosthesis depth using non-coronary cusp (NCcD) and left coronary cusp (LCcD) angiographic measurements.
Main Results:
- Higher rates of new atrioventricular block (AVB) and left ventricular bundle branch block (LBBB) were observed.
- 10% of patients required PPM implantation within 30 days.
- Significantly lower rates of conduction disturbances and PPM need were found with NCcD <6 mm or LCcD <8 mm (90-89%) compared to deeper implants (53-54%) (p <0.0001).
Conclusions:
- Prosthesis depth is a strong predictor of conduction disturbances after MCV TAVR.
- Angiographic measurement of prosthesis depth relative to the NCcD or LCcD can identify patients at higher risk.
- Optimizing implantation depth may reduce the need for PPM post-TAVR.
Related Concept Videos
Measurement of Blood Pressure
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Special considerations while measuring pulse
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Mitral Valve Prolapse II: Assessment and Management
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...