Variation in CT Number and Image Noise Uniformity According to Patient Positioning in MDCT
Timothy P Szczykutowicz1,2,3, Andrew DuPlissis3, Perry J Pickhardt1
11 Department of Radiology, University of Wisconsin-Madison, 1111 Highland Ave, Rm 1005, Madison, WI 53705.
Objective:
Many algorithms for clinical decision making rely on assessment of the CT number (expressed as Hounsfield units); however, to our knowledge, few, if any, studies have addressed how CT numbers change as a function of patient positioning within the scanner.
Materials And Methods:
An anthropomorphic phantom underwent imaging with varying amounts of vertical orientation misalignment with respect to isocenter. CT number and noise were measured using ROIs in the upper thorax, mid thorax, and abdomen. The degree of noise nonuniformity and changes in the CT number were assessed by comparing values obtained in the anterior versus posterior ROIs. To add clinical relevance, data on vertical mispositioning were collected from 20,316 clinical abdominal CT scans. Box-and-whisker plot analysis was used to identify the range of patient positioning.
Results:
Absolute CT number changes of more than 20 HU were observed for some ROIs at phantom positions of 10 cm from isocenter, with important differences noted between the thoracic and abdominal regions. Noise uniformity varied by more than twofold for all regions at 10 cm below isocenter. On clinical CT examinations, off-centering of more than 1, 2, 4, and 6 cm occurred for 41%, 19%, 1.9%, and 0.3% of patients, respectively.
Conclusion:
Radiologists should treat CT number measurements with caution when patients are grossly mispositioned in the scanner. The substantial changes in attenuation values shown in the present study are large enough to warrant further investigation.
More Related Videos
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
09:21Human Brown Adipose Tissue Depots Automatically Segmented by Positron Emission Tomography/Computed Tomography and Registered Magnetic Resonance Images
Published on: February 18, 2015
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...