Think Negative: Finding the Best Electrospray Ionization/MS Mode for Your Analyte
Piia Liigand1, Karl Kaupmees1, Kristjan Haav1
1Institute of Chemistry, University of Tartu , Ravila 14a, 50411, Tartu, Estonia.
A new method allows direct comparison of electrospray ionization efficiency (IE) in positive and negative modes. This reveals negative mode offers better sensitivity for nearly half of tested compounds, challenging common assumptions.
Area of Science:
- Analytical Chemistry
- Mass Spectrometry
- Separation Science
Background:
- Electrospray ionization (ESI) is a crucial technique in mass spectrometry for analyzing diverse compounds.
- Optimizing ionization conditions, specifically positive versus negative mode, is vital for maximizing sensitivity.
- Current methods lack a unified system for direct comparison of ionization efficiency (IE) across modes.
Purpose of the Study:
- To develop a novel system for directly comparing electrospray ionization efficiency (IE) in positive and negative modes.
- To establish a standardized approach for selecting optimal ionization conditions for enhanced analytical sensitivity.
- To challenge conventional practices by evaluating the sensitivity of positive versus negative ESI modes.
Main Methods:
- Development of a unified IE scale using a reference compound with similar ionization in both positive and negative modes.
- Analysis of 33 diverse compounds capable of ionizing in both positive and negative modes.
- Quantification and comparison of ionization efficiencies across both modes for each compound.
Main Results:
- A novel system successfully enabled direct comparison of IE values across positive and negative ionization modes.
- Negative mode provided superior sensitivity for 46% of the tested compounds, contrary to common assumptions.
- Positive mode was preferred for only 18% of compounds, while 36% showed comparable sensitivity in both modes.
Conclusions:
- The developed unified system effectively allows for the direct comparison of electrospray ionization efficiency across modes.
- Negative ion mode offers superior sensitivity for a significant portion of analytes, necessitating a re-evaluation of standard practices.
- This approach enables informed selection of optimal ionization conditions, enhancing analytical sensitivity and method development.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
07:46Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Chemical Ionization (CI) Mass Spectrometry
Mass Spectrum: Interpretation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Mass Spectrometry: Overview
Mass Analyzers: Overview