SWALLOWING COMPUTED TOMOGRAPHY: DOSE ESTIMATION IN A PHANTOM STUDY CONDUCTED AT VARIOUS PATIENT RECLINING ANGLES
Masanao Kobayashi1, Yasuki Asada1, Kosuke Matsubara2
1Graduate School of Health Sciences, Fujita Health University, Toyoake, Japan.
This study investigated patient radiation dose during swallowing computed tomography (SCT). The lowest effective dose was found at a 45° center reclining position, suggesting protocol optimization for safer swallowing assessments.
Area of Science:
- Medical Imaging
- Radiology
- Swallowing Disorders
Background:
- Swallowing computed tomography (SCT) is emerging for swallowing analysis.
- Optimal scan protocols for SCT are currently lacking.
- Patient radiation dose is a critical consideration in SCT.
Purpose of the Study:
- To estimate patient radiation dose at different reclining positions during SCT.
- To identify optimal scanning parameters for minimizing dose in SCT.
Main Methods:
- Utilized a RANDO phantom with thermoluminescent dosimeters.
- Simulated semi-reclining positions (45°, 55°, 65°) at center and off-center locations.
- Acquired scanograms following predetermined protocols to measure effective dose.
Main Results:
- The lowest effective dose (3.8 mSv) was observed at a 45° center reclining position.
- Off-center positions resulted in higher doses (4.6 mSv at 55°, 6.8 mSv at 65°) compared to center positions.
- Center positions yielded lower doses (4.5 mSv at 55°, 5.8 mSv at 65°) than off-center positions.
Conclusions:
- The 45° center reclining position is recommended for minimizing patient dose in SCT.
- Off-center scanning positions significantly increase radiation exposure and should be avoided.
- SCT protocols require revision to incorporate dose-optimization strategies based on patient position.
More Related Videos
08:32Adapting Human Videofluoroscopic Swallow Study Methods to Detect and Characterize Dysphagia in Murine Disease Models
Published on: March 1, 2015
06:33Construction of a Preclinical Multimodality Phantom Using Tissue-mimicking Materials for Quality Assurance in Tumor Size Measurement
Published on: July 29, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET