Ventricular Tachycardia After Implantable Cardioverter-Defibrillator Placement
Duy Thai Nguyen1, Zian H Tseng, Byron K Lee
1Cardiac Electrophysiology, Division of Cardiology, University of California, 500 Parnassus Avenue, MU East 4, Box 1354, San Francisco, CA 94143, USA.
Abstract:
Although implantation of an endocardial implantable cardioverter defibrillator (ICD) is meant to protect against ventricular arrhythmias, in some cases it can be paradoxically pro-arrhythmic. Recognition of this device complication, while rare, is important because it is potentially reversible and can be treated by managing the device, in lieu of, or in addition to, antiarrhythmics and catheter ablation of ventricular tachycardia (VT). This case describes VT caused by right ventricular pacing after ICD implantation.