Related Experiment Video
Updated: Jul 26, 2026

ampliPHOX Colorimetric Detection on a DNA Microarray for Influenza
Published on: June 9, 2011
A Phase-Intensity Surface Plasmon Resonance Biosensor for Avian Influenza A (H5N1) Detection
Chi Lok Wong1, Marissa Chua2, Heather Mittman3
1Bio-Optical Imaging Group, Singapore Bioimaging Consortium, Helios #01-02, 11 Biopolis Way, Singapore 138667, Singapore. dave_wong@sbic.a-star.edu.sg.
A novel phase-intensity surface plasmon resonance (SPR) biosensor offers sensitive, label-free detection of avian influenza A (H5N1) antibodies. This SPR biosensor significantly improves detection limits for influenza biomarker detection.
Area of Science:
- Biomedical Engineering
- Optoelectronics
- Biosensing Technology
Background:
- Surface Plasmon Resonance (SPR) is a label-free optical technique widely used for detecting biomolecular interactions.
- Traditional SPR methods can be complex, requiring interference fringe analysis or phase extraction.
- Avian Influenza A (H5N1) poses a significant public health threat, necessitating sensitive and rapid diagnostic tools.
Purpose of the Study:
- To develop and validate a novel phase-intensity SPR biosensor for sensitive avian influenza A (H5N1) antibody detection.
- To demonstrate the biosensor's capability for label-free, real-time detection without complex data processing.
- To compare the performance of the developed biosensor against existing commercial systems.
Main Methods:
- A phase-intensity SPR biosensor was designed utilizing the steep phase change at SPR excitation between p- and s-polarized light.
- The sensor head was integrated with polarizers to detect intensity variations corresponding to polarization ellipse rotation.
- Refractive index sensing and H5N1 antibody detection experiments were conducted to evaluate sensor resolution and sensitivity.
Main Results:
- The biosensor achieved a resolution of 6.3 × 10-6 refractive index units (RIU) in refractive index sensing.
- For H5N1 antibody detection, the sensor demonstrated a resolution of 193.3 ng mL-1.
- This represents a significant improvement in detection limit compared to commercial Biacore systems (517.3% and 258.7% better).
Conclusions:
- The developed phase-intensity SPR biosensor provides a sensitive, real-time, and label-free method for H5N1 antibody detection.
- The simplified detection mechanism avoids complex data analysis, making it user-friendly.
- This technology holds significant promise for improved influenza diagnostics and surveillance.
More Related Videos
07:15Preparation of Pseudo-Typed H5 Avian Influenza Viruses with Calcium Phosphate Transfection Method and Measurement of Antibody Neutralizing Activity
Published on: November 22, 2021
06:57In Vitro Culture for H5N1-Specific Duck T Cells and Detection of Immune Responses Using Intracellular Cytokine Staining Method
Published on: May 30, 2025