Reproducibility and Reliability of Repeated Quantitative Fluorescence Angiography
Nikolaj Nerup1, Kristine Bach Korsholm Knudsen2, Rikard Ambrus1
1Department of Surgical Gastroenterology, Copenhagen University Hospital Rigshospitalet, Copenhagen, Denmark.
Introduction:
When using fluorescence angiography (FA) in perioperative perfusion assessment, repeated measures with re-injections of fluorescent dye (ICG) may be required. However, repeated injections may cause saturation of dye in the tissue, exceeding the limit of fluorescence intensity that the camera can detect. As the emission of fluorescence is dependent of the excitatory light intensity, reduction of this may solve the problem. The aim of the present study was to investigate the reproducibility and reliability of repeated quantitative FA during a reduction of excitatory light.