Related Experiment Video
Updated: Feb 15, 2026

Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
In Vitro Validation of a Numerical Simulation of Leaflet Kinematics in a Polymeric Aortic Valve Under Physiological
Saleh Hassanzadeh Gharaie1, Bobak Mosadegh2, Yosry Morsi3
1Department of Radiology, Dalio Institute of Cardiovascular Imaging, New York Presbyterian Hospital & Weill Cornell Medicine, New York, NY, 10021, USA. shg2025@med.cornell.edu.
Abstract:
This paper describes a computational method to simulate the non-linear structural deformation of a polymeric aortic valve under physiological conditions. Arbitrary Lagrangian-Eulerian method is incorporated in the fluid-structure interaction simulation, and then validated by comparing the predicted kinematics of the valve's leaflets to in vitro measurements on a custom-made polymeric aortic valve. The predicted kinematics of the valve's leaflets was in good agreement with the experimental results with a maximum error of 15% in a single cardiac cycle. The fluid-structure interaction model presented in this study can simulate structural behaviour of a stented valve with flexible leaflets, providing insight into the haemodynamic performance of a polymeric aortic valve.
Related Concept Videos
Numerical Calculations
The solution to a problem is obtained using different methods. While manually solving algebraic symbols is one of the most common methods, the graphical method is often preferred. Computers...
Reliability and Validity
Kinematic Equations - I
Pilot and Numeric Relaying
How Data are Classified: Numerical Data
Quantitative data may be either discrete or continuous. All quantitative data that take on only specific numerical...
In Vitro Drug Release Testing: Overview, Development and Validation