EVALUATION OF PATIENT EXPOSURE IN FAST kVp SWITCHING DUAL ENERGY COMPUTED TOMOGRAPHY: PHANTOM STUDY
K Yagami1, T Miyoshi1, S Shigeyama1
1Radiology Services, Gifu University Hospital, Gifu, Japan.
Abstract:
The aim of this study was to evaluate the suitability of size specific dose estimates (SSDE) to estimate patient dose in Fast kVp switching dual energy CT. An anthropomorphic phantom (RAN-110) was repeatedly scanned (chest, abdomen and the pelvis) using a 64 detector row MDCT (Discovery CT750 HD, GE Healthcare, Milwaukee, WI, USA) with various CT parameters, including Fast kVp switching. Dosimetry was performed using thermo-luminescent dosimeters, positioned both superficially and within the phantom. SSDE was calculated for all slices of the anthropomorphic phantom using both the localiser and axial images. In Fast kVp switching, SSDE underestimated the measured absorbed dose for the chest/abdomen region ~35% at the maximum, but were in closer agreement for the pelvic region about within 10%. In single energy techniques, SSDE could not be applied in the estimation of organ doses, but in Fast kVp switching dual energy techniques, SSDE could be applied for anatomical regions with larger thicknesses.
More Related Videos
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Switching of BJT
Cut-off Mode ("Off" State): In this state, both the emitter-base and collector-base junctions are...
What is Energy?
Energy Basics
Free Energy