Testing of all Six Semicircular Canals with Video Head Impulse Test Systems
Published on: April 18, 2019
The Video Head Impulse Test and the Influence of Daily Use of Spectacles to Correct a Refractive Error
T S van Dooren1, F M P Lucieer1, A M L Janssen2,3
1Division of Balance Disorders, Department of Otorhinolaryngology and Head and Neck Surgery, Faculty of Health Medicine and Life Sciences, School for Mental Health and Neuroscience, Maastricht University Medical Centre, Maastricht, Netherlands.
Objective:
To determine the influence of daily use of spectacles to correct a refractive error, on the vestibulo-ocular reflex (VOR) gain measured with the video head impulse test (vHIT).
Study Design:
This prospective study enrolled subjects between 18 and 80 years old with and without a refractive error. Subjects were classified into three groups: (1) contact lenses, (2) spectacles, and (3) control group without visual impairment. Exclusion criteria comprised ophthalmic pathology, history of vestibular disorders, and alternated use of spectacles and contact lenses in daily life. Corrective spectacles were removed seconds before testing. One examiner performed all vHIT's under standardized circumstances using the EyeSeeCam system. This system calculated the horizontal VOR gain for rightward and leftward head rotations separately.
Results:
No statistically significant difference was found in VOR gain between the control group (n = 16), spectacles group (n = 48), and contact lenses group (n = 15) (p = 0.111). Both the spectacles group and contact lenses group showed no statistically significant correlation between VOR gain and amount of refractive error, for rightwards (p = 0.071) and leftwards (p = 0.716) head rotations. There was no statistical significant difference in VOR gain between testing monocularly or binocularly (p = 0.132) and between testing with or without wearing contact lenses (p = 0.800).
Conclusion:
In this study, VOR gain was not influenced by wearing corrective spectacles or contact lenses on a daily basis. Based on this study, no corrective measures are necessary when performing the vHIT on subjects with a refractive error, regardless of the way of correction.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Influence of Earth's Curvature and Atmospheric Refraction on Leveling
Impulse
Additionally, it can be shown that the...
Impulse Response
Kirchhoff's law forms an input signal equation, with the capacitor's current and voltage providing the output. Substituting the current and dividing by RC yields a differential equation. The output for an impulse input is the impulse...
Distance Corrections
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...