IAEA-SSDL bilateral comparisons for diagnostic level air kerma measurement standards
István Csete1, Paula Toroi1, Andreas Steuer2
1International Atomic Energy Agency (IAEA), P.O. Box 100, 1400 Vienna, Austria.
Abstract:
Comparisons of national standard of air kerma for conventional and mammographic diagnostic X-ray radiation qualities were conducted by the IAEA. Eleven secondary standards dosimetry laboratories provided calibration data for Exradin A3 and Radcal RC6M transfer ionization chambers circulated. Each comparison result expressed as the ratio of the participant and IAEA calibration coefficient were within the acceptance limit of ±2.5%. From the 67 results of 11 participants and 10 available beam qualities, the comparison result was within its standard uncertainty in 63 cases, and within the expanded (k = 2) uncertainty in four cases. Detailed calibration uncertainty budgets from participant laboratories are presented. The relative standard calibration uncertainty of each participant was in the range of 0.5-1.3%. These results indicate that the calibration related uncertainty component is reasonable low for a clinical measurement. In addition to the calibration coefficient, other corrections should be applied for clinical measurement to achieve the recommended accuracy.
Related Concept Videos
Measurement: Standard Units
Measurement of Air Content in Concrete
The pressure method,...
Ratio Level of Measurement
A set of data measured using the ratio scale takes care of the ratio problem and provides complete information. Ratio scale data are like interval scale data, except they have a zero point and ratios can be calculated....
Units and Standards of Measurement
Measurements of physical quantities are...
Ordinal Level of Measurement
Data measured using an ordinal scale are similar to nominal scale data, but there is one major difference. The ordinal scale data can be ordered. An example of ordinal scale data is a list of the top five national parks...
Interval Level of Measurement
Data measured using the interval scale are similar to ordinal level data because they have a definite arrangement. However, in the interval level of measurement, the differences between data values are meaningful even though the data does not have a starting point.
Temperature is measured using the interval scale. It is measurable data, and the difference between...