Full-root Aortic Valve Replacement by Stentless Aortic Xenografts in Patients with Small Aortic Roots
Published on: May 21, 2017
Outcomes of a New-Generation Stentless Aortic Valve: A Single-Center Experience with 251 Consecutive Implants
Guglielmo Stefanelli1, Fabrizio Pirro2, Giuseppe Danniballe2
1Hesperia Hospital, Modena, Italy. Electronic correspondence: guglielmostefanelli@gmail.com.
Background:
The study aim was to investigate the hemodynamic behavior over time and ease of implant and durability of the Sorin Pericarbon Freedom (SPF)® stentless pericardial aortic xenograft.
Methods:
Between March 2003 and April 2010, a total of 251 consecutive, non-selected patients (mean age 70.1 years; range: 17-89 years; 17.1% aged >80 years) received a SPF bioprosthesis as an aortic valve substitute at the authors' institution. All implantations were performed by a single surgeon using a classical, double-line, subcoronary implant technique. Of these patients. 108 (43%) underwent a concomitant procedure. The mean logistic EuroSCORE was 8.3. Patients were followed for complications and hemodynamic evaluation. Echocardiographic controls and clinical data were obtained at discharge, and at six months' and eight years' follow up.
Results:
The in-hospital/30-day mortality was 1.2% for the entire group, but 0% for patients with isolated valve replacement. A total of 27 deaths had occurred at the time of the last follow up (22 were due to non-cardiac causes). At follow up (mean 3.7 years; range: 0-7.8 years), 91.1% of survivors were in NYHA classes I or II. Freedom from reoperation and from structural valve deterioration was 96.0% and 96.8%, respectively, at 7.8 years. The mean pressure gradient of the series (measured using echocardiography) was 10.3 ± 4.5 mmHg at discharge, and 8.3 ± 4.5mmHg at the time of the last follow up. The mean effective orifice area was 1.85 ± 0.70 cm2 for the entire series, and there was an absence of prosthesis-patient mismatch in 90% of patients.
Conclusions:
The study results confirmed the favourable hemodynamic behavior, low risk of implantation and acceptable durability of the pericardial stentless SPF aortic prosthesis. The data also endorsed use of the SPF as an aortic valve substitute, even in a young and active population, and particularly in the presence of a small aortic annulus. The accuracy of implantation remains a crucial factor for durability and subsequent satisfactory hemodynamic performance. A longer follow up would provide further information with regards to complications and durability of the SPF prosthesis.
Related Concept Videos
Lattice Centering and Coordination Number
Types of Unit Cells
Imagine taking a large number of identical...
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Predicting Reaction Outcomes
Center of Gravity
Center of Gravity
To determine its location, the principle of moments can be utilized by dividing the object into...
Outcomes of Glycolysis
Cellular respiration can occur aerobically (with oxygen) or anaerobically (without oxygen). In the presence of oxygen, cellular respiration starts with glycolysis and continues with pyruvate...