Gripping and Anchoring Effects on the Mechanical Strengths of Orthodontic Microimplants
Chun-Ming Chen1, Chun-Chan Ting2, Huang-Chi Wang3
1Professor, Department of Oral and Maxillofacial Surgery, Kaohsiung Medical University Hospital, Kaohsiung, Taiwan.
Purpose:
The aim of this study was to evaluate the mechanical strengths in 5 different designs of orthodontic microimplants by analyzing their configuration of structure.
Materials And Methods:
Thirty microimplants of 5 types (diameter 1.5 mm: type A, B, and C; diameter 1.3 mm: type D and E) were assessed. All microimplants were manually driven into the artificial bones at a 7-mm depth. The anchor area (AA), gripping area (GA), insertion torque (IT), Periotest value (PTV), and pullout strength (PS) were measured. Intergroup and intragroup comparisons were used to detect their significant differences.
Results:
In the intergroup comparison, type D had a least IT (4.5 Ncm). In the PTV analysis, type B had the largest AA (7.76 mm) and its PTV (1.6) was significantly least than the others. In the PS test, type C had the largest GA (2.40 mm) and its PS was the largest. Intragroup comparisons (IT and PS), type A, and type E presented positively significant correlation. GA revealed positive with PS, and AA showed reverse tendency with PTV.
Conclusion:
The more AA of microimplants, the more stable they are. The more GA of microimplants, the more PS they are. Therefore, type C was better than the others because it had the largest GA and second largest AA.
Related Concept Videos
Ionic Strength: Effects on Chemical Equilibria
In this solution, the primary...
The Anchoring-and-Adjustment Heuristic
Anchoring Junctions
Lipids as Anchors
The carboxy-terminal of most of the prenylated proteins, such as Ras proteins, contains...
Acid Strength and Molecular Structure
In the absence of any leveling effect, the acid strength of binary compounds of hydrogen with nonmetals (A) increases as the H-A bond strength decreases down a group in the periodic table. For group 17, the order of increasing acidity is HF < HCl < HBr < HI. Likewise, for group 16, the order of increasing acid strength is H2O < H2S < H2Se < H2Te. Across a row in the periodic table, the acid strength of binary hydrogen compounds increases with increasing...
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...