External electrical cardioversion in patients with cardiac implantable electronic devices: Is it safe and is
Nikki A H A Pluymaekers1, Elton A M P Dudink1, Lucas Boersma2
1Department of Cardiology, Cardiovascular Research Institute Maastricht (CARIM), Maastricht University Medical Center (MUMC+), Maastricht, The Netherlands.
Background:
Atrial tachyarrhythmias are common in patients with cardiac implantable electronic devices (CIEDs). Restoration of sinus rhythm by external electrical cardioversion (eECV) is frequently used to alleviate symptoms and to ensure optimal device function.
Objectives:
To evaluate the safety of eECV in patients with contemporary CIEDs and to assess the need for immediate device interrogation after eECV.
Methods:
We conducted a retrospective observational study of 229 patients (27.9% female, age 69 ± 10 years) with a CIED (104 pacemakers, 69 implantable cardioverter defibrillators, and 56 biventricular devices) who underwent eECV between 2008 and 2016 in two centers. Data from device interrogation before eECV, immediately afterwards, and at first follow-up (FU) after eECV were collected. CIED-related complications and adverse events during and after eECV were recorded.
Results:
No significant differences between right atrial (RA) and right ventricular (RV) sensing or threshold values before eECV, immediately afterwards, or at FU were observed. A small yet significant decrease was observed in RA and RV impedance immediately after eECV (484 Ω vs 462 Ω, P < 0.001 and 536 Ω vs 514 Ω, P < 0.001, respectively). The RV impedance did not recover to the baseline value (538 Ω vs 527 Ω, P = 0.02). The impedance changes were without clinical consequences. No changes in left ventricular lead threshold or impedance values were measured. No CIED-related complications or adverse events were documented following eECV.
Conclusion:
eECV in patients with contemporary CIEDs is safe. There seems to be no need for immediate device interrogation after eECV.
Related Concept Videos
Non-ohmic Devices
Consider a simple circuit consisting of a battery, a diode, and a resistor. A...
Mnemonic Devices
Acronyms
Acronyms are created by using the initial letters of a series of words to form a new word or phrase. This approach condenses complex information into a single, memorable entity. For example,...
Work Done on a System by External Force
In the presence of a non-conservative opposing force, like friction, some part of the work done...
Electron Behavior
Electrons are negatively charged subatomic particles that are attracted to an orbit around the positively-charged nucleus of an atom. They reside in locations that are associated with energy levels called shells and are further organized into sub-shells and orbitals within each shell.
Electrons Orbit the Nucleus
Electrons are found in specific locations outside of the nucleus. The shell in which an electron resides indicates the general energy level of the electron: those closer to the...
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device
Guidelines and Strategies for Safe Computer Charting
Maintain Confidentiality and Security: