A Spine Robotic-Assisted Navigation System for Pedicle Screw Placement
Published on: May 11, 2020
Development of a shoulder-mounted robot for MRI-guided needle placement: phantom study
Reza Monfaredi1, Iulian Iordachita2, Emmanuel Wilson3
1Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Health System, Washington, DC, USA. RMonfare@childrensnational.org.
The Arthrobot, an MRI-guided needle placement robot, demonstrated minimal image distortion and noise. Its targeting accuracy is sufficient for shoulder arthrography, with a combined average error of 3.4 mm.
Area of Science:
- Medical robotics
- Image-guided interventions
- Orthopedic procedures
Background:
- MRI-guided interventions require precise instrument navigation.
- Existing systems may face challenges with image quality and accuracy.
- Arthrobot is a novel patient-mounted robot designed for MRI-guided needle placement.
Purpose of the Study:
- To quantitatively evaluate the performance of the Arthrobot system.
- To assess signal-to-noise ratio (SNR), image distortion, and targeting accuracy.
- To determine the suitability of Arthrobot for procedures like shoulder arthrography.
Main Methods:
- Designed the robot and its inverse kinematics.
- Conducted SNR and distortion studies using phantoms.
- Evaluated mounting stability with a respiratory motion study.
- Performed phantom studies for end-to-end targeting accuracy assessment.
Main Results:
- SNR changes were minimal (2%) with the robot present.
- Maximum image distortion was 2.57%.
- Average targeting error was 1.64 mm (SD 0.90 mm), with motion contributing 1.32 mm (SD 1.38 mm).
Conclusions:
- Arthrobot minimally impacts MRI image quality (SNR and distortion).
- The system achieves satisfactory targeting accuracy for MRI-guided procedures.
- Combined average total error is 3.4 mm (SD 1.65 mm), meeting clinical requirements for shoulder arthrography.
Related Concept Videos
Muscles of the Shoulder
Anterior Thoracic Muscles
The anterior thoracic muscles include the serratus anterior, subclavius, and...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Sustainable Development
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Case Studies
Development of the Lymphatic System
The first lymph sacs to form are the paired jugular lymph sacs located at the junction of the internal jugular and subclavian veins. From these sacs, lymphatic capillary plexuses extend to the thorax, upper limbs, neck, and head, eventually forming lymphatic vessels. Each jugular lymph sac maintains a...