Multicenter study of quantitative PET system harmonization using NIST-traceable 68Ge/68Ga cross-calibration kit
Kenta Miwa1, Kei Wagatsuma2, Takashi Iimori3
1Department of Radiological Sciences, School of Health Science, International University of Health and Welfare, Tochigi, Japan.
This study harmonized Standardized Uptake Values (SUV) in PET/CT scanners using a NIST-traceable 68Ge/68Ga source. The novel method reduced SUV variability to within 5% across sites, improving multicenter study reliability.
Area of Science:
- Nuclear Medicine
- Medical Imaging
- Radiochemistry
Background:
- Standardized Uptake Value (SUV) is crucial for quantitative analysis in Positron Emission Tomography/Computed Tomography (PET/CT).
- Variability in SUV measurements across different PET/CT scanners can hinder multicenter studies and clinical decision-making.
- Harmonization of SUV is essential for consistent and reliable quantitative PET imaging.
Purpose of the Study:
- To define errors in SUV measurements.
- To demonstrate the feasibility of harmonizing SUV across contemporary PET/CT scanners.
- To establish a novel reference standard using a National Institute of Standards and Technology (NIST)-traceable 68Ge/68Ga source.
Main Methods:
- Utilized a NIST-traceable 68Ge/68Ga source for dose calibrator and PET source preparation.
- Determined bias in radioactivity and concentration measurements at five Japanese sites.
- Optimized dose calibrator settings and PET reconstruction parameters for SUV harmonization.
Main Results:
- Average bias in radioactivity and concentration measurements reduced from -4.94% and -12.22% to -0.14% and -4.81% post-harmonization.
- Clinical SUV errors improved from -7.66% underestimation to -4.70% under optimal conditions.
- SUV variability was reduced to within 5% across different PET scanner models.
Conclusions:
- The proposed method effectively identified and corrected bias in dose calibrator and PET scanner measurements.
- NIST-traceable 68Ge/68Ga sources provide a reliable standard for SUV harmonization.
- This protocol shows significant potential for standardizing SUV in multicenter PET/CT studies.
Related Concept Videos
Harmonic Mean
Take the example of the speed of a car, which is the measure of the rate of distance traveled. If the vehicle traverses the same distance back-and-forth, its average speed equals the total distance traveled divided by the total time taken. However, if the car moves with varying speeds, then the arithmetic mean is more skewed...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Crossing Over
The homologous pairs of sister chromosomes—one from the maternal and one from the paternal genome—then begin to align alongside each other lengthwise, matching corresponding DNA positions in a process...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Cross-Sectional Research
Monohybrid Crosses