iPhone-based Pupillometry: A Novel Approach for Assessing the Pupillary Light Reflex
J Jason McAnany1,2, Brandon M Smith3, Amy Garland4
1Department of Ophthalmology and Visual Sciences, University of Illinois at Chicago, Chicago, Illinois.
Significance:
The response of the pupil to a flash of light, the pupillary light reflex (PLR), is an important measure in optometry and in other fields of medicine that is typically evaluated by qualitative observation. Here we describe a simple, portable, iPhone-based pupillometer that quantifies the PLR in real time.
Purposes:
The purposes of this study were to describe a novel application that records the PLR and to compare its technical capabilities with a laboratory-based infrared (IR) camera system.
Methods:
Pupil sizes were measured from 15 visually normal subjects (age, 19 to 65 years) using an IR camera system and the Sensitometer test. This test elicits pupillary constriction using the iPhone flash, records pupil size using the camera, and provides measurements in real time. Simultaneous recordings were obtained with the Sensitometer test and IR camera, and two measures were calculated: (1) dark-adapted steady-state pupil size and (2) minimum pupil size after the flash. The PLR was defined as the difference between these two measures. Pupil size was also recorded during the redilation phase after the flash. Bland-Altman analysis was used to assess the limits of agreement between the two methods.
Results:
Statistically significant correlations between the IR and Sensitometer test measures were found for the PLR (r = 0.91, P < .001) and redilation size (r = 0.65, P = .03). Bland-Altman analysis indicated a mean PLR difference of 6% between these two methods. The PLR limit of agreement was 14%, indicating that 95% of subjects are expected to have IR and Sensitometer test measurements that differ by 14% or less. Bland-Altman analysis indicated a mean redilation size difference of 1% between the two methods; the limit of agreement was 5%.
Conclusions:
There is excellent agreement between pupil responses recorded using the Sensitometer test and IR camera. The Sensitometer test provides a highly promising approach for simple, portable, inexpensive pupillary measurements.
Related Concept Videos
Reflex Activity
A reflex exam is a diagnostic procedure performed by a healthcare professional to evaluate the functionality of a patient's...
The Micturition Reflex
The process begins with bladder filling, where the bladder wall stretches as urine accumulates. This stretching activates the urine storage reflex, mediated by the sacral spinal segments and the pontine storage center. Efferent sympathetic impulses stimulate the detrusor muscle to relax and the internal urethral sphincter to contract, facilitating...
Somatic Spinal Reflexes
One of the most well-known somatic spinal reflexes is the stretch reflex, which is activated by the sudden stretching of a muscle. This reflex involves the activation of specialized sensory receptors called muscle spindles, which are located in the muscle tissue and detect changes in the length and speed of muscle contractions. When a muscle is suddenly...
Light Acquisition
Light as Energy
Photons
A photon is a discrete electromagnetic particle or bundle of energy. Photons are characterized by their frequency, wavelength, and amplitude, similar to the properties of a wave. Waves with higher frequencies transmit more energy and have shorter wavelengths than longer wavelengths that transmit...
The Wave Nature of Light