Precision in 3-Dimensional Surface Imaging of the Face: A Handheld Scanner Comparison Performed in a Cadaveric Model
Konstantin C Koban1, Sebastian Cotofana2, Konstantin Frank1
1Department for Hand, Plastic and Aesthetic Surgery, Ludwig - Maximilian University, Munich, Germany.
Background:
Handheld 3-dimensional surface imaging (3DSI) devices of various precision are becoming more versatile in their applications and more widely accepted by clinicians for documentation.
Objectives:
The authors sought to compare the precision of facial volumetric change measurements of 3 3DSI devices in the cadaveric model: Eva (Artec 3D Inc., Luxembourg), Sense (3D Systems, Rock Hill, SC), and iSense (3D Systems, Rock Hill, SC).
Methods:
A total of 336 scanning and analysis procedures were carried out in 4 cephalic specimens (mean age, 77.25 ± 24.3 years; mean BMI, 21.76 ± 6.6 kg/m2). Two superficial and 2 supraperiosteal regions of interest were injected with 0.5-cc aliquots and subsequently scanned using the 3 different scanners. Correlation coefficients between the injected and measured volume were computed.
Results:
The correlation coefficient for the Eva scanner was for subcutaneous regions of interest rp = 0.935 and for the supraperiosteal regions of interest rp = 0.966, compared with rp = 0.760 and rp = 0.364 (superficial vs supraperiosteal) for the Sense and rp = 0.694 and rp = 0.382 (superficial vs supraperiosteal) for the iSense scanner.
Conclusions:
3DSI devices are capable of measuring surface volume changes of the face at a level of 0.5-cc surface volume change and can thus be regarded as useful tools in the preinterventional, intrainterventional, and postinterventional phases of a treatment. One of the 3 evaluated scanners provided very high correlation coefficients between the injected and the measured volume (Eva), whereas the other evaluated 3DSI devices provided moderate (Sense) and low (iSense) coefficients.
More Related Videos
07:57Scaled Anatomical Model Creation of Biomedical Tomographic Imaging Data and Associated Labels for Subsequent Sub-surface Laser Engraving SSLE of Glass Crystals
Published on: April 25, 2017
06:48Author Spotlight: Advancements in 3D Optical Imaging for Comprehensive Body Composition Assessment in Modern Research
Published on: June 7, 2024
Related Concept Videos
Pharmacokinetic Models: Comparison and Selection Criterion
Physiological models take a detailed approach by considering specific molecular processes. They can predict drug distribution, metabolism, and elimination changes, providing a comprehensive understanding of how drugs interact with the body.
The Sense of Self: Reflected Self-Appraisal and Social Comparison
Dimensional Analysis
Conversion Factors and Dimensional Analysis
The unit...
Uncertainty in Measurement: Accuracy and Precision
Molecular Comparison of Gases, Liquids, and Solids
Multiple Comparison Tests
It would be easy to compare two samples using a significance alpha level of 0.05. In other words, there is only one sample pair to be compared. However, it would be difficult to identify a significantly different sample if the number...