Selected Reaction Monitoring Mass Spectrometry for Absolute Protein Quantification
Published on: August 17, 2015
Isotope dilution LC-ESI-MS/MS and low resolution selected reaction monitoring as a tool for the accurate
Amaia Ereño Artabe1, Adriana González-Gago1, Amanda Suarez Fernández1
1Department of Physical and Analytical Chemistry, Faculty of Chemistry, University of Oviedo, Spain.
Abstract:
A new analytical method for the quantification of testosterone in human urine samples by isotope dilution mass spectrometry is proposed. A standard solution of 13C2-testosterone is added to the samples at the beginning of the sample preparation procedure and then the measurements are carried out by UHPLC-ESI-MS/MS. In the proposed method, the resolution of the first quadrupole of the tandem MS instrument is reduced to transmit the whole precursor ion cluster to the collision cell and measure the isotopic distribution of the in-cell product ions with a small number of SRM transitions. The construction of a methodological calibration graph is avoided using a labelled analogue previously characterised in terms of concentration and isotopic enrichment in combination with multiple linear regression. In this way, the molar fractions of natural and labelled testosterone are calculated in each sample injection and the amount of endogenous testosterone computed from the known amount of labelled analogue. Recovery values between 97 and 107% and precisions between 0.4 and 3.7% (as %RSD) were obtained for testosterone concentrations in urine in the range of 1 to 8 ng g-1. The proposed low resolution SRM methodology was compared for the analysis of human urine samples with the traditional IDMS method based on a calibration graph and the IDMS method based on multiple linear regression combined with standard resolution SRM. A similar accuracy and precision was obtained by the three tested approaches. However, using the low resolution SRM method there was no need to resort to calibration graphs or to specific dedicated software to calculate isotopic distributions by tandem MS and a higher sensitivity was obtained. The proposed low resolution SRM method was successfully applied to the analysis of the certified freeze-dried human urine NMIA MX005.
Related Concept Videos
Isotopes
An element's atomic mass, or weight,...
Elements: Chemical Symbols and Isotopes
Some symbols are derived from the common English name of the element; others are abbreviations of the name in another language — Latin, Greek or German. For example, the symbol for aluminum (common name)...
Testosterone: Functions and Regulation
Solution Concentration and Dilution
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Half-life of a Reaction