Reliability and Validation of the Hexoskin Wearable Bio-Collection Device During Walking Conditions
Jeffrey Montes1, John C Young1, Richard Tandy1
1Department of Kinesiology and Nutrition Sciences, University of Nevada, Las Vegas, Las Vegas, NV, USA.
International Journal of Exercise Science
|October 20, 2018
Summary
The Hexoskin smart shirt accurately measures heart rate and respiratory rate during walking. However, minute ventilation, step count, and energy expenditure measurements varied in accuracy depending on walking speed.
Area of Science:
- Sports Medicine
- Wearable Technology
- Physiological Monitoring
Background:
- Smart shirts offer continuous physiological monitoring for athletes and medical applications.
- Validating wearable sensor accuracy against established methods is crucial for reliable data collection.
Purpose of the Study:
- To assess the validity and reliability of the Hexoskin smart shirt (HxS) for measuring heart rate (HR), respiratory rate (RR), minute ventilation (VE), step count (SC), and energy expenditure (EE).
- To compare HxS measurements against gold-standard equipment during a controlled walking protocol.
Main Methods:
- A two-day walking treadmill protocol was conducted with 49 participants at speeds of 1.5, 2.5, and 3.5 mph.
- HxS data was compared to a Polar T-31 heart rate monitor, an Applied Electrochemistry Moxus Metabolic System, and manual step counts.
- Validity was assessed using Pearson's correlation (r ≥ 0.70), and reliability using Cronbach's alpha (α ≥ 0.70).
Main Results:
- HxS demonstrated valid and reliable measurements for HR and RR across all tested speeds.
- VE, SC, and EE showed mixed results, with accuracy varying significantly based on walking speed.
- Specifically, HxS HR showed high correlation (r=0.86) and reliability (α=0.86) at 1.5 mph, while EE showed moderate correlation (r=0.56) and reliability (α=0.85) at the same speed.
Conclusions:
- The Hexoskin smart shirt is a valid and reliable tool for measuring heart rate and respiratory rate during walking.
- Further research is needed to improve the accuracy of VE, SC, and EE measurements at different speeds for broader applications.
- These findings are significant for optimizing the use of Hexoskin in athletic training, coaching, and telemedicine.
More Related Videos
Related Concept Videos
Reliability and Validity
13.9K
Reliability and validity are two important considerations that must be made with any type of data collection. Reliability refers to the ability to consistently produce a given result. In the context of psychological research, this would mean that any instruments or tools used to collect data do so in consistent, reproducible ways.
13.9K
Distribution Reliability and Automation
515
Distribution reliability in electrical power systems is critical for ensuring an uninterrupted power supply to consumers at minimal cost. According to IEEE Standard Terms, reliability is the probability that a device will function without failure over a specified time period or amount of usage. For electric power distribution, this translates to maintaining continuous power supply and addressing customer concerns over power outages. Several indices, as defined by IEEE Standard 1366-2012, are...
515
Data Validation
1.8K
Method validation is a crucial process in analytical chemistry designed to confirm that a given method consistently produces reliable and high-quality results. This process is essential when a method is applied to different sample matrices or when procedural modifications are made, ensuring that the results meet acceptable standards across various applications.
Key parameters for method validation include:
Key parameters for method validation include:
1.8K
Data Validation
6.8K
Data validation is an essential part of a comprehensive assessment. Validation is confirming or verifying and opening the door to gathering more assessment data as it clarifies vague or unclear data. The process of checking and verifying the collected information is called data validation. The primary purpose of data validation is to ensure data is as free from error, bias, and misinterpretation as possible.
Nursing assessment guides are generally based on holistic models rather than medical...
Nursing assessment guides are generally based on holistic models rather than medical...
6.8K
Data Collection II
10.1K
The nursing history captures and records the patient's health status, so that a care plan evolves to meet the patient's individual needs. The nursing health history is a part of the initial assessment. A comprehensive history covers all health dimensions and plays a significant role in the assessment process. A comprehensive history includes the patient's biographical information, reasons for seeking health care, expectations, present and past health history, medications, and...
10.1K
Formulating and Validating Nursing Diagnosis I
3.9K
A nursing diagnosis is written when the nurse recognizes a cluster of essential patient data indicating health problems treated with independent nursing interventions. The standardized terminologies of a nursing diagnosis help nurses identify and treat patients' problems. Every electronic health record that uses nursing diagnosis must employ standard diagnostic terminology. Developing an efficient, individualized care plan begins with accurate nursing diagnoses.
There are thirteen domains...
There are thirteen domains...
3.9K