Related Experiment Video
Updated: Feb 3, 2026

Applying Dynamic Strain on Thin Oxide Films Immobilized on a Pseudoelastic Nickel-Titanium Alloy
Published on: July 28, 2020
Mechanical Properties of Orifice Preflaring Nickel-titanium Rotary Instrument Heat Treated Using T-Wire Technology
Maamoun Ataya1, Jung-Hong Ha2, Sang Won Kwak3
1Endodontic Department, Hamdan Bin Mohammed College of Dental Medicine, Mohammed Bin Rashid University of Medicine and Health Sciences, Dubai, United Arab Emirates.
Introduction:
This study examined whether the use of T-Wire heat treatment enhanced the resistance to torsional force, cyclic fatigue, and bending stiffness of orifice preflaring nickel-titanium instruments.
Methods:
Ninety heat-treated OneFlare (OFT; Micro-Mega, Besançon, France) and conventional OneFlare (OFC, Micro-Mega) nickel-titanium files of size #25/.09 with identical geometries were selected (n = 15 per group for each test). Torsional resistance was evaluated through the measurement of ultimate torsional strength, distortion angle, and toughness during rotational loading. Cyclic fatigue resistance was evaluated through measurement of the number of cycles to failure for each instrument. Stiffness was evaluated by observation of the bending moment on attaining a 45° bend. Independent t tests were performed for statistical analysis. After torsional and cyclic fatigue tests, all fractured fragments were examined under a field emission scanning electron microscope to observe characteristics of the fractured surfaces.
Results:
OFT showed better cyclic fatigue resistance and lower bending stiffness than OFC (P < .05); however, OFT and OFC did not differ significantly in terms of torsional resistance. The fractured cross-sectional surfaces had characteristics typical of cyclic fatigue and torsional fractures.
Conclusions:
The OFT instruments made using T-Wire heat treatment had better cyclic fatigue resistance and lower bending stiffness with no decline in torsional strength.
Related Concept Videos
Mechanisms of Heat Transfer
Conduction, accounting for approximately 3% of body heat loss at rest, is the process of exchanging heat between molecules of two materials in direct contact. This can result in both heat loss and gain. For instance, when the body is submerged in water, which conducts heat 20 times more effectively than air, it can either lose or gain significant...
Mechanism of heat transfer
Mechanisms of Heat Transfer I
Mechanisms of Heat Transfer II
Quantifying Heat
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...