Background:
Cone beam CT (CBCT) imaging has been integrated into the most recent version of the Leksell Gamma Knife for the primary purpose to facilitate fractionated therapy.
Case Description:
This case study presents three patients where the CBCT system of the Gamma Knife Icon discovered potentially clinically significant frame shifts. In each case, patients were imaged with volumetric MR prior to stereotactic frame placement. Immediately following frame placement, diagnostic stereotactic CT imaging was acquired with a stereotactic indicator box attached to the frame. Following treatment planning and immediately before radiosurgery, a CBCT was acquired using the on-board imaging functionality of the Gamma Knife Icon, which provides a registration of the patient's anatomy to stereotactic space independent of that provided by the stereotactic frame/fiducials. Co-registration of the CT and CBCT provides an estimate of the difference between these two estimates of stereotactic coordinates. The vector magnitudes of the differences measured at the center of stereotactic space were 0.93mm, 2.64mm and 2.18 mm for Case 1, Case 2 and Case 3 respectively.
Conclusions:
Use of the CBCT functionality of the Gamma Knife Icon to verify the consistency of frame placement can prevent clinically significant targeting errors due to frame slippage or frame adapter mounting errors, and allows any required adjustments to be made without interrupting the overall treatment workflow.
Related Concept Videos
Quality Assurance
Framing Effects
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Frames
Frames are versatile and widely used in various applications such as structural supports for beams and columns, automobile chassis construction, and in the construction...
Imaging Studies for Cardiovascular System V: CT
Frames: Problem Solving II