Related Experiment Video
Updated: Jan 30, 2026

Fabrication of Uniform Nanoscale Cavities via Silicon Direct Wafer Bonding
Published on: January 9, 2014
In Vitro Study of Bond Strength Between Abutments with Different Degrees of Convergence and Crowns by Pre-Bonding
Purpose:
To investigate the optimal degree of convergence of the abutment with which the bond strength achieved by the pre-bonding method is comparable with that in direct bonding with a conventional degree of convergence.
Materials And Methods:
Abutments with 5.5-mm diameter, 5-mm height, 0.5-mm shoulder width, and three kinds of degrees of convergence (2, 4, and 6 degrees) were first designed by digital modeling. Their corresponding inner crowns were also modeled, and a gap of 40 μm was kept between the abutment and the inner crown. Thirty abutments and 30 inner crowns were then lathed out from a titanium plate (10 sets per degree of convergence). Six groups were defined in this study, according to the different degrees of convergence and bonding methods (direct bonding, pre-bonding) (n = 10 sets). The samples handled with direct bonding would be cleaned for reuse in tests with pre-bonding. Temporary cement was used as an adhesive, and the bond strength was tested in each set of samples. The comparison among the results was performed by the Kruskal-Wallis test.
Results:
The mean values of bond strength with direct bonding methods were 349.39 ± 65.75 N, 316.49 ± 54.22 N, and 277.49 ± 56.96 N, and with pre-bonding methods were 279.35 ± 48.58 N, 227.97 ± 26.72 N, and 154.6 ± 23.03 N, respectively (2, 4, and 6 degrees). No statistical difference was found among the values in direct bonding groups and, in pre-bonding groups, only the comparison between 2 and 6 degrees of convergence showed statistical significance (P = .000). Between different bonding methods, statistical differences were shown in abutments with 4 and 6 degrees of convergence (P = .006, P = .000), respectively. The bond strength with pre-bonding methods and 2 degrees of convergence showed no significant difference from that with direct bonding and 6 degrees of convergence.
Conclusion:
The bond strength was inversely proportional to the degree of convergence, and the bond strength of pre-bonding was lower than that of direct bonding with the same degree of convergence. When using the pre-bonding method, the bond strength between the abutment and inner crown with 2 degrees of convergence could be comparable with using the direct bonding method and abutments with conventional degrees of convergence.
Related Concept Videos
Bonding and Strength of Aggregate
Bond Energies and Bond Lengths
Ionic Bonds
When atoms gain or lose electrons to achieve a more stable electron configuration they form ions. Ionic bonds are electrostatic attractions between ions with opposite charges. Ionic compounds are rigid and brittle when solid and may dissociate into their constituent ions in water. Covalent compounds, by contrast, remain intact unless a chemical reaction breaks them.
Opposing Charges Hold Ions Together in Ionic Compounds
Ionic bonds are reversible electrostatic interactions between ions...
Peptide Bonds
Bonding in Metals
Valence Bond Theory