Related Experiment Video
Updated: Jan 29, 2026

Optimizing Minimally Invasive Spine Surgery: A Fully 3D CT O-Arm Navigated Workflow in MIS TLIF
Published on: October 17, 2025
RADIATION EXPOSURE DURING SPINE SURGERY USING C-ARM FLUOROSCOPY
Alexandre Fogaça Cristante1, Fábio Barbieri2, Almy Anacleto Rodrigues da Silva2,3
1Universidade de São Paulo, Faculdade de Medicina, Hospital das Clínicas (IOT-HCFMUSP), Instituto de Ortopedia e Traumatologia, Spine Surgery Division, Laboratory of Medical Investigation, São Paulo, SP, Brazil.
Objective:
To evaluate the radiation dose received by staff in spine surgeries, including those who are not considered occupationally exposed workers.
Methods:
All spinal surgeries performed in the same department during a period of 12 months were evaluated with regard to the exposure of surgeons, scrub nurses, and auxiliary personnel working in the operating room to radiation from C-arm fluoroscopy. Radiation was measured by 15 film badge dosimeters placed on the professionals' lapels, gloves, and room standardized sites. The films were analyzed in the dosimetry laboratory by collections per period.
Results:
During the 12 months, 81 spinal surgeries were performed by the same team, with surgical times ranging from 1 to 6 hours. The total radiation dose ranged from 0.16 mSv to 2.29 mSv depending on the dosimetry site. The most exposed site was the wrist of the main surgeon.
Conclusion:
The results showed that in the spinal surgeries in our setting, the radiation doses are low and within legal limits. Nevertheless, constant training of professionals is essential for radiation protection of medical staff and patients.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Muscles that Move the Arm
The pectoralis major has two origins. Its clavicular head originates on the medial half of the clavicle. In contrast, the sternocostal head originates on the costal cartilages of ribs 1-6, the sternum, and the aponeurosis of the external oblique of the...
Radiation: Applications
The average...
Absorption of Radiation
Radiation Pressure: Problem Solving
The average value of the rate of momentum transfer divided by the absorbing area represents the average force...
Generating Electromagnetic Radiations