Modal analysis of two different types of fixed implant-supported prostheses embedded in edentulous maxillae
Akihiro Suzuki1, Taiji Hoshiai2, Hidemi Nakata1
1Department of Oral Implantology and Regenerative Dental Medicine, Graduate School of Medical and Dental Sciences, Tokyo Medical and Dental University, Tokyo, Japan.
Purpose:
Modal analysis has been used to monitor and quantify the vibration of each component of a prosthesis at its natural frequency. An understanding of the vibrational characteristics of a prosthesis can guide selection of the appropriate prosthesis configuration to prevent excessive stress. In this study, modal analysis of fixed implant-supported prostheses with 4 or 6 abutments in edentulous maxillae was performed.
Methods:
Sixteen patients underwent maxillary dental implant treatment and received screw-fixed implant-supported prostheses (9 patients received 4-abutment prostheses and 7 patients received 6-abutment prostheses) in edentulous maxillae. An impact hammer was used for excitation of the prostheses, and vibration was detected on every crown of the prostheses. The modal parameters were determined in each subject. Furthermore, the modal shapes were compared at each natural frequency.
Results:
The median natural frequencies of subjects with 4- and 6-abutment prostheses were 697 and 781Hz, respectively. The mean damping ratios did not differ significantly between 4- (8.2±1.7%) and 6-abutment (6.6±1.7%) prostheses (p=0.125). The mean maximum displacements of fixed prostheses were significantly lower for 6-abutment (10.3±3.3μm) than for 4- (20.3±9.1μm) prostheses (p=0.004). The median number of vectors in antiphase was significantly lower for 6- (4) than for 4-abutment prostheses (14) (p=0.001).
Conclusions:
The current study demonstrated less adverse vibration in the 6-abutment prostheses than in the 4-abutment prostheses, suggesting that modal analysis can contribute to novel future developments in the designs of dental implant prostheses.
Related Concept Videos
Sensory Modalities
General senses refer to the broad category of sensory information detected by receptors in the body and can be further grouped into somatic and visceral senses. Somatic sensations include touch, pressure, temperature, and pain and are essential for navigating our environment and...
Self-Help Support Groups
Accessibility and Cost-Effectiveness
One of the primary strengths of self-help...
Fixing Double-strand Breaks
Fixed Action Patterns
Support Reactions
The purpose of the supports is to prevent the translational motion of the system by applying an equal and opposite force and to prevent the system's rotation by applying...
Classification of Titrimetric Analysis Based on Reaction Types
Titrations between an acid and a base lead to neutralization reactions that form...