A Rare Cause of Stapedotomy Failure: When the Depth Marker Becomes a Locking Device: A Report on 2 Cases
Mary Daval1, Nicolas Guevara2, Naif Alotaibi1
11 Department of Otorhinolaryngology Head and Neck Surgery, Fondation Ophtalmologique Adolphe de Rothschild, Paris, France.
Abstract:
We report 2 cases of stapes surgery failure due to the depth marker groove of the piston of a Big Easy titanium prosthesis getting caught at the edge of the stapedotomy hole. In the first case, the stapedotomy failure, detected peroperatively, was corrected by modifying the angulation of the shaft of the prosthesis. In the second case, the failure occurred after a delay during which scar tissue developed around the depth marker groove, hampering the movement of the prosthesis. The usual postoperative imaging did not detect this complication. The diagnosis was made during revision surgery and the condition was treated by enlarging the hole in the footplate. Alternatively, the use of a piston without a depth marker groove would have resolved the problem. Although a rare event, the possibility of stapes surgery failure due to the depth marker groove should not be overlooked.
More Related Videos
Related Concept Videos
Self-Locking Screw
A square-threaded screw jack carrying a load is considered self-locking if the screw retains its position even after the moment applied to it is removed.
Uniform Depth Channel Flow
Depth Perception and Spatial Vision
Data Reporting and Recording
Uniform Depth Channel Flow: Problem Solving
Types of Reports I: Hands-off Report
Following are the key components and categories of hand-off reports:
Purpose and Process: