Detection of high-frequency artifact as a function of pulse generator algorithms and outer-insulation material
Ryohsuke Narui1, Tomofumi Nakamura1, Ikutaro Nakajima1
1Arrhythmia Section, Cardiovascular Division, Department of Medicine, Vanderbilt University Medical Center, Nashville, Tennessee.
Background:
A high rate of malfunction, predominantly intermittent high-frequency artifacts (HFAs), has been recently reported in Abbott Medical Tendril pacing leads.
Objective:
To investigate the factors associated with the occurrence of HFAs on Tendril leads using a commonly used comparator lead for a control.
Methods:
We reviewed institutional data for Medtronic CapSureFix 5076 and Abbott Medical Tendril pace-sense leads retrospectively. Recordings deemed to be due to electromagnetic interference and far-field oversensing were not included in the classification of HFAs.
Results:
A total of 7673 leads were analyzed: 1628 Optim-insulated Tendril leads, 825 non-Optim Tendril leads, and 5220 CapSureFix 5076. HFAs were seen in 212 leads and were more frequently observed in Tendril compared to CapSureFix leads during a mean follow-up of 4.1 ± 3.6 years. Lower age at implant, defibrillator systems, atrial position, and connection to an Abbott Medical generator were associated with increased HFA. In multivariable analysis, only connection to Abbott Medical generators (odds ratio 7.686, P < .001) and age (odds ratio 0.988 per year, P = .016) were independently associated with HFAs on pace-sense leads. In an Abbott-generator-only analysis, Optim-insulated Tendril leads were more likely to display HFAs than non-Optim Tendril leads but not Medtronic CapSureFix 5076 leads.
Conclusion:
Abbott Medical pulse generators independently predict HFA in Tendril and CapSureFix 5076 leads, likely the result of displaying short or low-amplitude noise episodes that other devices do not record. When restricted to Abbott generators only, Optim-insulated Tendril leads show an increased incidence of HFAs when compared to non-Optim Tendril leads but not CapSureFix 5076 leads.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
10:23Equipment Setup and Artifact Removal for Simultaneous Electroencephalogram and Functional Magnetic Resonance Imaging for Clinical Review in Epilepsy
Published on: June 23, 2023
Related Concept Videos
Rectangular and Triangular Pulse Function
For example, consider a rectangular pulse with a 5V amplitude, a 3-second duration, and centered at t=2 seconds. This pulse can be expressed using the rectangular function, written as,
Insulation Coordination
Conductors and Insulators
Most metals are conductors. Their atomic configuration is such that one or more electron(s) are loosely bound to the nucleus in each atom. Thus, a sea of mobile electrons are available in them, known as free electrons. Their easy...
Thermal Insulation in Masonry Walls
External insulation can be applied using an Exterior Insulation and Finish System (EIFS), which involves affixing panels of plastic foam to the wall and covering them with a polymeric stucco reinforced with glass fiber mesh....
Pulse
The pulse serves as a clinical...
Pulse
Pulse Rate and its Significance
Pulse rate, often measured in beats per minute (bpm), reflects the heart rate (HR), which is influenced by numerous factors such as stress, physical activity, and hormonal changes. A normal resting adult pulse rate falls...