A semi-quantitative rapid multi-range gradient lateral flow immunoassay for procalcitonin.
Kseniya V Serebrennikova1,2, Jeanne V Samsonova3,4, Alexander P Osipov3,4
1Chemistry Faculty, Lomonosov Moscow State University, Moscow, Leninskiye Gory, 119991, Russia. ksenijasereb@mail.ru.
A new gradient lateral flow immunoassay (LFIA) offers rapid, semi-quantitative procalcitonin (PCT) detection for early sepsis diagnosis. This visual test strip method effectively identifies PCT concentration ranges, aiding in timely clinical decisions for severe bacterial infections.
Area of Science:
- Biomedical Engineering
- Immunodiagnostics
- Clinical Chemistry
Background:
- Procalcitonin (PCT) is a crucial biomarker for diagnosing sepsis and bacterial infections.
- Existing diagnostic methods may lack the speed or semi-quantitative capability needed for rapid clinical assessment.
- Lateral flow immunoassays (LFIA) offer potential for rapid point-of-care diagnostics.
Purpose of the Study:
- To develop a rapid, semi-quantitative gradient lateral flow immunoassay (LFIA) for procalcitonin (PCT).
- To establish a visual method for determining PCT concentration ranges relevant to sepsis diagnosis.
- To evaluate the performance of the developed LFIA against established methods like ELISA.
Main Methods:
- A gradient LFIA was designed by immobilizing antibodies at increasing concentrations on test strips.
- Two types of gold nanoparticles (spherical and popcorn-like) were used as labels.
- Visual interpretation of colored test lines correlated to specific PCT concentration ranges (˂0.25 to ≥10 ng·mL⁻¹).
Main Results:
- The gradient LFIA successfully provided semi-quantitative analysis of PCT levels across defined ranges.
- Comparison with ELISA (n=82) showed high sensitivity (87.5%) and specificity (92.3%) at key PCT thresholds.
- The assay demonstrated reliable performance for identifying clinically relevant PCT concentration ranges.
Conclusions:
- The developed multi-range gradient LFIA is a promising tool for the early diagnosis of sepsis and severe bacterial infections.
- This visual, semi-quantitative method offers a practical approach for rapid PCT assessment.
- The gradient LFIA technology has broad applicability for other analytes in various LFIA-based diagnostic systems.
Related Concept Videos
Rapidly Varying Flow
What is an Electrochemical Gradient?
The chemical gradient relies on differences in the abundance of a substance on the outside versus the inside of a cell and flows from areas of high to low ion concentration. In contrast, the electrical gradient revolves around an...
Range
15.9; 16.1; 15.2; 14.8; 15.8; 15.9; 16.0; 15.5
Measurements of the amount of soda in a 16-ounce can vary since different subjects record these measurements or since the exact amount - 16 ounces of liquid, was not...
¹H NMR: Long-Range Coupling
In alkenes, spin information is communicated via σ–π overlap, as seen in allylic (four-bond) and homoallylic (five-bond) couplings. These coupling interactions are stronger when the σ bond is parallel to the alkene...
Lateralization
Variation: Normal Distribution, Range, and Standard Deviation